文献速递乳房整形与再造显微修复重建

带磁端口乳房组织扩张器在1.5T磁共振血管造影中的安全性

Breast Tissue Expanders with Magnetic Ports: Clinical Experience at 1.5 T.
PubMed 文献 · Plast Reconstr Surg · 2016-01-01

该回顾性研究评估了71例携带标注为「MR Unsafe」的磁端口乳房组织扩张器患者,在自体乳房重建前于1.5T下行腹部/盆腔磁共振血管造影的安全性。结果显示无扩张器或磁端口移位、无扫描疼痛、无组织损伤及手术并发症,认为该检查可安全实施。

要点

  • 磁端口扩张器在1.5T磁共振血管造影中未发生移位或疼痛。
  • 扩张器取出时未发现组织损伤或重建部位手术并发症。
  • 该检查可用于自体皮瓣乳房重建前的穿支血管定位。

核心结论

  • 携带特定磁端口乳房组织扩张器的患者可在1.5T下安全进行腹部/盆腔磁共振血管造影,用于自体皮瓣乳房重建前穿支血管定位。

研究设计与样本

  • 回顾性研究,2012年7月至2014年5月纳入71例携带标注「MR Unsafe」磁端口组织扩张器的患者,在1.5T下俯卧位行腹部/盆腔磁共振血管造影。

主要发现

  • 磁共振检查期间无患者出现组织扩张器或磁端口移位,无患者报告扫描时疼痛。
  • 取出组织扩张器时(71例患者,112个植入物),术者报告无组织损伤证据,乳房重建部位无手术并发症。

临床怎么用

  • 对携带特定磁端口乳房组织扩张器的患者,可在1.5T下行腹部/盆腔磁共振血管造影进行穿支血管定位,以辅助自体皮瓣乳房重建规划。

局限性

  • 证据等级为治疗性IV级,为回顾性研究。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The purpose of this study was to evaluate breast tissue expanders with magnetic ports for safety in patients undergoing abdominal/pelvic magnetic resonance angiography before autologous breast reconstruction. METHODS: Magnetic resonance angiography of the abdomen and pelvis at 1.5 T was performed in 71 patients in prone position with tissue expanders with magnetic ports labeled "MR Unsafe" from July of 2012 to May of 2014. Patients were monitored during magnetic resonance angiography for tissue expander-related symptoms, and the chest wall tissue adjacent to the tissue expander was examined for injury at the time of tissue expander removal for breast reconstruction. Retrospective review of these patients' clinical records was performed. T2-weighted fast spin echo, steady-state free precession and gadolinium-enhanced spoiled gradient echo sequences were assessed for image artifacts. RESULTS: No patient had tissue expander or magnetic port migration during the magnetic resonance examination and none reported pain during scanning. On tissue expander removal (71 patients, 112 implants), the surgeons reported no evidence of tissue damage, and there were no operative complications at those sites of breast reconstruction. CONCLUSION: Magnetic resonance angiography of the abdomen and pelvis in patients with certain breast tissue expanders containing magnetic ports can be performed safely at 1.5 T for pre-autologous flap breast reconstruction perforator vessel mapping. CLINICAL QUESTION/LEVEL OF EVIDENCE: Therapeutic, IV.

来源:Plast Reconstr Surg,查看原文。版权归原作者所有。

引用

  1. Thimmappa ND, Prince MR, Colen KL, et al. Breast Tissue Expanders with Magnetic Ports: Clinical Experience at 1.5 T. Plast Reconstr Surg. 2016. PMID: 27879583. doi:10.1097/PRS.0000000000002782