三维成像系统在DIEP皮瓣乳房再造修整术中的体积评估

Volumetric assessment using a three-dimensional imaging system in revision of breast reconstruction with deep inferior epigastric artery perforator flaps.
PubMed 文献 · J Plast Reconstr Aesthet Surg · 2026-05-07

本研究回顾性分析VECTRA® H1三维成像在DIEP皮瓣乳房再造后修整减容手术中的体积评估准确性。术前及术后6个月拍摄三维图像,比较术后体积差与实际切除组织重量的-致性。结果显示ICC(2,1)为0.654、ICC(2,k)为0.791(p<0.0001),Bland-Altman偏倚2.28,一致性界限-85.4至89.96,简单回归R²=0.43(p<0.001),提示三维体积差与实际切除重量一致,可作为自体组织乳房再造修整术的体积参考指标。

要点

  • VECTRA® H1测得的双侧乳房体积差与实际切除组织重量具有一致性。
  • ICC(2,1)=0.654、ICC(2,k)=0.791,提示中等至良好信度。
  • VECTRA® H1可作为DIEP皮瓣乳房再造后修整减容术的体积参考工具。

核心结论

  • VECTRA® H1评估的双侧乳房体积差与实际切除组织重量一致,可作为自体组织乳房再造后修整术的有用体积指标。

研究设计与样本

  • 回顾性研究,纳入DIEP皮瓣乳房再造后行减容修整术的患者,术前及术后6个月行VECTRA® H1三维成像,分析术后体积差与实际切除组织重量的一致性。

主要发现

  • ICC(2,1)和ICC(2,k)分别为0.654和0.791(p<0.0001)。
  • Bland-Altman图偏倚为2.28,一致性界限为-85.4至89.96。
  • 简单回归分析显示显著正相关(y=0.547x+34.72,R²=0.43,p<0.001)。

临床怎么用

  • 在DIEP皮瓣乳房再造后修整减容术中,可参考VECTRA® H1三维成像测得的体积差估计需切除的组织量。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Three-dimensional imaging systems are used in breast volume revision surgery. However, the utility of a three-dimensional imaging system for breast reduction after autologous breast reconstruction is yet to be reported. In this study, we retrospectively investigated the use of VECTRA® H1 (Canfield Scientific, Inc. NJ. USA.) three-dimensional image analysis to estimate the amount of tissue required for breast revision surgery after deep inferior epigastric artery perforator flap reconstruction. METHODS: This retrospective study included patients who underwent reconstructive breast volume reduction surgery after reconstruction with a deep inferior epigastric artery perforator flap. We took three-dimensional images before and 6 months after surgery, and analyzed the consistency between the volume difference (apparent difference) after surgery and weight of the actual reduced tissue. RESULTS: The intraclass correlation coefficient [ICC(2,1) and ICC(2,k)] were 0.654 and 0.791 (p < 0.0001), respectively. The Bland-Altman plot yielded a bias of 2.28, with limits of agreement ranging from -85.4 to 89.96. A significant positive correlation was found between these values in a simple regression analysis (y = 0.547x + 34.72, R2 = 0.43, p < 0.001). CONCLUSION: To our knowledge, this is the first study examining the accuracy of VECTRA® H1 for tissue reduction, including skin and fat. The difference in breast volume between the two sides assessed using VECTRA® H1 was consistent with the actual resection weights. Therefore, VECTRA® H1 could be a useful volumetric indicator for revision surgery after breast reconstruction using autologous tissue.