微吸引背景装置在大鼠股血管模型显微吻合中的应用
作者研发并获专利一种三维打印微流体背景装置,可自主持续冲洗和吸引,在大鼠股动脉模型上由两名资深显微外科医生完成12例端端吻合(传统组6例、吸引辅助背景SAB组6例)。SAB组吻合时间更短、SAMS评分更高、擦拭事件更少,均达统计学显著。
要点
- 大鼠股动脉端端吻合中,SAB组完成时间短于传统组(584±155.9秒 vs 741.7±203.1秒,P=0.007)。
- SAB组SAMS评分高于传统组(38.3±1.5 vs 32.3±1.4,P=0.001)。
- SAB组擦拭事件中位数少于传统组(1.7±1.2 vs 13±2.2,P<0.001)。
核心结论
- 新型微流体背景装置可无需助手实现持续冲洗和吸引,优化显微血管吻合效率。
研究设计与样本
- 研究在大鼠股血管模型中进行,由两名资深显微外科医生完成12例端端吻合,传统组6例、吸引辅助背景(SAB)组6例。
- 主要结局为完成吻合所需时间;次要结局包括经验证的结构化显微外科技能评估(SAMS)评分和获得术野清晰所需的擦拭事件总数。
- 每例操作录像,由两名盲法专家独立使用SAMS评分评定。
主要发现
- 完成吻合时间传统组为741.7±203.1秒,SAB组为584±155.9秒,传统组大于SAB组(P=0.007)。
- SAMS评分传统组为32.3±1.4,SAB组为38.3±1.5,传统组低于SAB组(P=0.001)。
- 擦拭事件中位数传统组为13±2.2,SAB组为1.7±1.2,传统组显著多于SAB组(P<0.001)。
临床怎么用
- 作者设计的新型获专利三维打印显微外科背景装置可提供持续冲洗和吸引,与标准背景相比减少手术时间并在显微吻合中提供更好的血管清晰度。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
SUMMARY: Microvascular anastomoses can be challenging to perform when edematous fluids and blood continuously flood and compromise the field of view. Intermittent irrigation and suctioning disturb workflow, require an assistant, and can increase risk of arterial thrombosis from vessels being drawn into suction drains. The authors developed and patented a novel three-dimensionally printed background device with microfluidic capabilities to provide autonomous, continuous irrigation and suction to optimize operator autonomy and efficiency. The authors tested this in a rat femoral vessel model. Twelve end-to-end anastomoses were performed by two senior microsurgeons [six conventional, six suction-assisted background (SAB)] in a rat femoral artery model. The primary outcome was time taken to complete the anastomosis. Secondary outcomes included the validated Structured Assessment of Microsurgery Skills (SAMS) score and the total number of "wiping" events to obtain field clarity. Each procedure was recorded, and videos were independently rated by two blinded experts using the SAMS score. Time taken to complete the anastomosis was greater in the conventional group compared with the SAB group (741.7 ± 203.1 seconds versus 584 ± 155.9 seconds; P = 0.007). The median SAMS score was lower in the conventional group compared with the SAB group (32.3 ± 1.4 versus 38.3 ± 1.5; P = 0.001). The median number of wiping events was significantly greater in the conventional group compared with the SAB group (13 ± 2.2 versus 1.7 ± 1.2; P < 0.001). The authors show that a novel microfluidic background device allows continuous irrigation and suctioning without the need for an assistant, optimizing the efficiency of the microvascular anastomosis. CLINICAL RELEVANCE STATEMENT: The authors have designed a novel, patented, three-dimensionally printed microsurgical background device that provides continuous irrigation and suction, reduces operative time, and provides better vessel clarity during a microsurgical anastomosis compared to standard background.
引用
- Khajuria A, Jeong HH, Papavasiliou T, et al. Application of a Microsuction Background Device for Microanastomosis in a Rat Femoral Vessel Model. Plast Reconstr Surg. 2023. PMID: 37014957. doi:10.1097/PRS.0000000000010512
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。