增强现实虚拟投影对微血管穿支血管空间保真度的感知评估
本研究利用19例患者CT血管造影数据制作腹壁下深动脉穿支皮瓣相关解剖的三维打印模型,并由4名独立观察者从两个视点比较模型与微软HoloLens全息投影上对应解剖点的空间偏差。结果显示,三维模型与视点1之间的中位平移偏差为9.0 mm,三维模型与视点2之间为12.1 mm,视点1与视点2之间为13.5 mm;作者认为对应点可在毫米级识别,但误差仍受观察者和投影感知位置等多个独立因素影响。
要点
- 三维模型与视点1投影对应点的中位平移偏差为9.0 mm。
- 三维模型与视点2投影对应点的中位平移偏差为12.1 mm。
- 视点1与视点2投影对应点的中位平移偏差为13.5 mm。
核心结论
- 用于乳房再造的穿支血管地形等对应点可在毫米级范围内被识别,但仍存在多个独立误差来源,最显著的是观察者及感知投影的位置。
研究设计与样本
- 利用19例去标识患者CT血管造影数据制作腹壁下深动脉穿支皮瓣相关解剖的三维打印模型,并由4名独立观察者比较三维模型与其HoloLens虚拟投影之间的空间保真度。
主要发现
- 所有观察者中,三维模型与视点1对应点的中位平移偏差为9.0 mm。
- 所有观察者中,三维模型与视点2对应点的中位平移偏差为12.1 mm。
- 所有观察者中,视点1与视点2对应点的中位平移偏差为13.5 mm。
- 采用广义线性模型等统计方法评估全息投影在平移、旋转和尺度方面的空间保真度。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Autologous breast reconstruction yields improved long-term aesthetic results but requires increased resources of practitioners and hospital systems. Innovations in radiographic imaging have been increasingly used to improve the efficiency and success of free flap harvest. Augmented reality affords the opportunity to superimpose relevant imaging on a surgeon's native field of view, potentially facilitating dissection of anatomically variable structures. To validate the spatial fidelity of augmented reality projections of deep inferior epigastric perforator flap-relevant anatomy, comparisons of three-dimensional (3D) models and their virtual renderings were performed by four independent observers. Measured discrepancies between the real and holographic models were evaluated. METHODS: The 3D-printed models of deep inferior epigastric perforator flap-relevant anatomy were fabricated from computed tomographic angiography data from 19 de-identified patients. The corresponding computed tomographic angiography data were similarly formatted for the Microsoft HoloLens to generate corresponding projections. Anatomic points were initially measured on 3D models, after which the corresponding points were measured on the HoloLens projections from two separate vantage points (V1 and V2). Statistical analyses, including generalized linear modeling, were performed to characterize spatial fidelity regarding translation, rotation, and scale of holographic projections. RESULTS: Among all participants, the median translational displacement at corresponding points was 9.0 mm between the real-3D model and V1, 12.1 mm between the 3D model and V2, and 13.5 mm between V1 and V2. CONCLUSION: Corresponding points, including topography of perforating vessels, for the purposes of breast reconstruction can be identified within millimeters, but there remain multiple independent contributors of error, most notably the participant and location at which the projection is perceived.
引用
- Cholok DJ, Fischer MJ, Leuze CW, et al. Spatial Fidelity of Microvascular Perforating Vessels as Perceived by Augmented Reality Virtual Projections. Plast Reconstr Surg. 2023. PMID: 37092985. doi:10.1097/PRS.0000000000010592
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。