三维技术测量瘢痕表面积的效度与信度的临床计量学评估

A clinimetric assessment of the validity and reliability of 3D technology for scar surface area measurement.
PubMed 文献 · Burns · 2022-12-23

本研究评估了移动设备深度传感器三维系统测量瘢痕表面积的效度和信度。效度研究在20名健康志愿者体表放置80个已知面积贴纸(8.01 cm²至77.70 cm²),观察者A的变异系数为0.62%至1.67%,观察者B为0.75%至1.19%。信度研究纳入36例患者的69处瘢痕,观察者间组内相关系数为0.99至1.00,校正观察者间变异系数为5.59%。

要点

  • 深度传感器三维系统测量瘢痕表面积具有效度,贴纸面积测量变异系数低于1.67%。
  • 观察者间信度极高,组内相关系数为0.99至1.00。

核心结论

  • 带有深度传感器的三维技术测量瘢痕表面积提供有效且可靠的数据,从而超越了昂贵且耗时的三维相机。

研究设计与样本

  • 效度研究:在20名健康志愿者的不同身体部位放置80个预定尺寸贴纸(8.01 cm²至77.70 cm²),单观察者设置下成像。
  • 信度研究:纳入36例患者的69处瘢痕(肥大性瘢痕、瘢痕疙瘩和正常营养性瘢痕),由两名观察者独立成像和评级。
  • 使用移动设备上安装的深度传感器,通过定制软件应用进行图像采集和分析,遵循标准化单关键帧图像采集程序。

主要发现

  • 效度研究:观察者A的变异系数为0.62%至1.67%,观察者B为0.75%至1.19%。
  • 信度研究:平均瘢痕表面积为13.83 cm²(标准差23.06)对观察者A,13.59 cm²(标准差23.31)对观察者B。
  • 观察者间组内相关系数为0.99至1.00。
  • 校正观察者间变异系数估计为5.59%,对应一致性界限为0±0.15倍平均表面积。

关键数值与推荐

  • 效度以变异系数评估,观察者A为0.62%至1.67%,观察者B为0.75%至1.19%。
  • 信度以组内相关系数和一致性界限评估,观察者间组内相关系数为0.99至1.00,校正观察者间变异系数为5.59%,一致性界限为0±0.15倍平均表面积。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

INTRODUCTION: The quality of scars has become an important outcome of burn care. Objective scar assessment through scar surface area measurement enables quantification of scar formation and evaluation of treatment efficacy. 3D technology has proven valid and reliable but often remains cumbersome, expensive, and time-consuming. 3D technology with depth sensors on mobile devices has become available and might surpass these limitations. This study provides a clinimetric assessment of the validity and reliability of a 3D system with a depth sensor for scar surface area measurement. METHODS: A technology involving a depth sensor mounted on a mobile device was used. Images and analyses were made with a custom-made software application. A standardized one-keyframe image capturing procedure was followed. To assess validity, stickers with predefined dimensions (8.01 cm2 - 77.70 cm2) were imaged in a single observer setting on various body parts of healthy volunteers. To assess reliability, hypertrophic scars, keloids, and normotrophic scars were imaged and rated by two observers independently. Data are expressed as mean (+/-SD), Coefficient of Variation (CV), Intraclass Correlation Coefficients (ICC), and Limits of Agreements (LoA). RESULTS: Eighty stickers placed on 20 healthy volunteers showed validity with CV between 0.62%- 1.67% for observer A and 0.75%- 1.19% for observer B. For the reliability study, 69 scars on 36 patients were included. Mean scar surface area ranged from 0.83 cm2 to 155.59 cm2. Mean scar surface area measurement was 13.83 cm2 (SD 23.06) for observer A and 13.59 cm2 (SD 23.31) for observer B. Adjusted interobserver CV for trained observers is estimated as 5.59%, with corresponding LoA = 0 ± 0.15 x mean surface area. Interobserver ICCs were 0.99-1.00. CONCLUSION: This 3D technology with a depth sensor for measuring scar surface area provides valid and reliable data and thereby surpasses expensive and time-consuming 3D cameras.