临床实践中使用Cutometer®进行客观烧伤瘢痕评估:标准化测量方案的介绍与验证
本研究开发了一套适用于日常临床实践的Cutometer® MPA 580标准化测量方案,包含基于解剖标志的瘢痕定位技术,并通过4名临床医生对17例患者34处烧伤瘢痕的评估验证其评分者间信度。参数R0(瘢痕柔韧性)平均测量ICC为0.92(95%CI 0.86;0.96),单次测量ICC为0.74(0.61;0.84);测量探头施加的压力被识别为影响可靠测量的混杂因素,评分者性别不影响测量信度。
要点
- 标准化Cutometer®方案对烧伤瘢痕R0参数平均测量ICC 0.92(95%CI 0.86;0.96),单次测量ICC 0.74(0.61;0.84)。
- 测量探头施加的压力被识别为可靠测量的影响性混杂因素。
- 评分者性别不影响测量信度。
核心结论
- 所引入的Cutometer® MPA 580标准化测量方案可用于临床及研究目的的客观、可靠烧伤瘢痕评估。
研究设计与样本
- 开发Cutometer® MPA 580标准化测量方案,包括基于解剖标志的瘢痕定位技术,经历识别关键因素、初步测试、技术细节确定、方案细化及最终测试等步骤。
- 通过4名临床医生对17例患者的34处烧伤瘢痕进行评估,计算组内相关系数(ICC)以验证评分者间信度。
主要发现
- 参数R0代表瘢痕柔韧性,被确定为最合适的输出参数,平均测量评分者间信度极好(ICC 0.92 [95% CI 0.86; 0.96]),单次测量信度可接受(ICC 0.74 [0.61; 0.84])。
- 测量探头施加的压力被识别为影响可靠测量的混杂因素。
- 评分者性别不影响测量信度。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
An objective burn scar assessment is essential to informed therapeutic decision-making and to monitor scar development over time. However, widely employed scar rating scales show poor inter-rater reliability. For this study we developed a standardized measurement protocol for the Cutometer© applicable for objective burn scar assessment in everyday clinical practice. We developed a measurement protocol for the Cutometer© MPA 580 including a scar site relocation technique based on anatomical landmarks. The protocol emerged through several steps: Identifying key factors for valid and reliable measurements, preliminary testing, specification of technical details, refining the protocol and final testing. Consecutively, the protocol was validated for inter-rater reliability by assessing 34 burn scars in 17 patients by four clinicians and computing an Intra-class Correlation Coefficient (ICC). Parameter R0, representing scar pliability, was identified as the best suited output parameter yielding excellent inter-rater reliability for average measures (ICC 0.92 [95% CI 0.86; 0.96]) and acceptable reliability for single measures (ICC: 0.74 [0.61; 0.84]). The pressure applied on the measuring probe was identified as an influential confounding factor for reliable measurements. Rater gender did not influence reliability of measurements. The introduced standardized measurement protocol for the Cutometer© MPA 580 enables an objective and reliable burn scar assessment for clinical as well as research purposes.
引用
- Klimitz FJ, Neubauer H, Stolle A, et al. Objective Burn Scar Assessment in Clinical Practice Using the Cutometer©: Introduction and Validation of a Standardized Measurement Protocol. J Burn Care Res. 2023. PMID: 36300728. doi:10.1093/jbcr/irac154
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。