鼻测量角器:一种鼻整形术用测量仪器的验证
研究在31例初次鼻整形患者术中测量下外侧软骨间发散角,将鼻测量角器与计算机分析对比,并评估两名术者间的可重复性。两种方法测得值存在显著差异(p=0.034),角器平均小3.7度;角度≤75度时角器平均高0.79度,不同术者间差异显著(p<0.023)。
要点
- 31例初次鼻整形患者术中测量下外侧软骨发散角,鼻测量角器与计算机分析对比差异显著(p=0.034)。
- 角器测得值平均比软件小3.7度;角度≤75度时角器平均高0.79度,>75度时两法差异更大(p=0.022)。
- 不同术者使用鼻测量角器的测量差异显著(p<0.023),但作者结论称其可重复。
核心结论
- 角度≤75度时,鼻测量角器比软件测量平均高0.79度,差异可忽略;鼻测量角器在不同术者间显示可重复。
研究设计与样本
- 在31例初次鼻整形患者术中测量下外侧软骨两支外侧支间的发散角,将鼻测量角器与计算机分析对比,并评估两名不同术者使用角器的可重复性。
主要发现
- 鼻测量角器与计算机分析测值比较差异显著(p=0.034),角器平均比软件小3.7度。
- 两法测角均>75度时,两法差异更大(p=0.022);角度≤75度时,角器平均比软件高0.79度。
- 不同术者使用鼻测量角器测得的角度差异显著(p<0.023)。
关键数值与推荐
- 下外侧软骨头侧偏位时,软骨间发散角通常小于60度。
临床怎么用
- 原文结论称鼻测量角器在角度≤75度时与软件差异可忽略,且在不同术者间可重复。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
The position of the lower lateral cartilages is directly related to good nasal functionality. When these cartilages exhibit cephalic malposition, the angle of divergence between cartilages is usually less than 60 degrees, which can cause external nasal valve insufficiency. The objective of this study was to validate the rhinogoniometer, an innovative surgical instrument that allows for diagnosing the position of this cartilage intraoperatively. In the intraoperative period, the angle of divergence, which is the angle formed between the two lateral branches of the lower lateral cartilages, was measured in 31 primary rhinoplasty patients. The rhinogoniometer measurement was compared with that of the computerized analysis, and the reproducibility of measurements with the rhinogoniometer between two different surgeons was determined. When comparing the values obtained by the two methods, a significant difference was found (p = 0.034). On average, the angles measured by the rhinogoniometer were 3.7 degrees smaller than the measurements made by the software. It was observed that when the angles measured by both methods were above 75 degrees, the differences between the two measurements were higher (p = 0.022). However, for angles up to 75 degrees, the measurement using the rhinogoniometer was, on average, 0.79 degrees higher than the measurement taken by the software. When the difference in angles obtained by the two methods between different surgeons was evaluated, there was a significant difference (p < 0.023). In conclusion, at angles up to 75 degrees, the rhinogoniometer showed a negligible difference of 0.79 degrees higher than the measurement obtained via software. Measurements with the rhinogoniometer were also shown to be reproducible with different surgeons.
引用
- Silva EN, Dos Santos TK, Ferreira LM, et al. Rhinogoniometer: Validation of an Instrument for Rhinoplasty. Plast Reconstr Surg. 2021. PMID: 34847112. doi:10.1097/PRS.0000000000008587
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。