药物类型与输注时长对静脉智能泵报警的影响
结果发现抗菌药物产生更高比例的「门打开」和「管路气泡」报警,阿片激动剂产生更多患者自控镇痛相关报警,抗肿瘤药和单克隆抗体有高报警/输注比。按药物类别评估IVSP报警数据可识别减少不必要报警的靶点。
要点
- 抗菌药物输注中「门打开」和「管路气泡」报警比例高于全部输注样本。
- 阿片激动剂产生更多患者自控镇痛相关报警,中枢神经系统药物「患者侧阻塞」报警较少。
- 按药物类别或单个药物分析IVSP报警数据可定位需改进的报警问题区域。
核心结论
- 按药物类别或单个药物评估IVSP报警数据可用于识别减少不必要报警的潜在靶点。
主要发现
- 抗菌药物输注产生的「门打开」和「管路气泡」报警比例高于全部输注样本。
- 中枢神经系统药物「患者侧阻塞」报警较少,阿片激动剂产生更多患者自控镇痛相关报警。
- 抗肿瘤药、单克隆抗体及血液形成、凝血和血栓形成药物均具有高的报警/输注比(有报警的输注)。
- 抗肿瘤药和单克隆抗体产生高比例管路气泡报警,血液形成、凝血和血栓形成药物(如肝素)有更多患者侧阻塞报警。
临床怎么用
- 临床可依据药物类别或具体药物分析IVSP报警数据,定位高频报警类型并制定针对性改进措施。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: Alarm fatigue is a serious problem in hospitals, often affecting nurses' concentration and increasing the risk of patient harm. Alarms associated with intravenous smart pumps (IVSPs) contribute to alarm fatigue; however, little is known about what causes unnecessary IVSP alarms or how to reduce them. The purpose of this study is to explore variables that impact the type and number of IVSP alarms and to demonstrate how alarm data can be used to identify IVSP alarm problem areas to target for improvement. METHODS: A dataset containing IVSP infusion and alarm data for 62 million infusions from 107 US hospitals over a 24-month period was analyzed. Detailed analyses on the types and frequency of IVSP alarms for medication classes and select medications were performed. RESULTS: Over 65 million alarms were analyzed. Infusions of antibacterials generated a higher percentage of "Door Open" and "Bolus Air in Line" alarms than were generated in the all-infusion sample. Central nervous system (CNS) agents had fewer "Patient Side Occlusion" alarms, and opioid agonists generated more patient-controlled analgesia-related alarms. Antineoplastic agents, monoclonal antibodies, and agents for blood formation, coagulation, and thrombosis all had a high ratio of alarms to infusions with alarms. Antineoplastic agents and monoclonal antibodies generated a high percentage of Bolus Air in Line alarms, and agents for blood formation, coagulation, and thrombosis (ie, heparin) had more Patient Side Occlusion alarms. CONCLUSION: This study demonstrates how IVSP alarm data, when evaluated by medication class or individual medication(s), can be used to identify potential areas to target for reducing unnecessary IVSP alarms.
引用
- Armistead LT, Giuliano KK, Degnan D, et al. Exploring the impact of medication type and infusion duration on IV smart pump alarms. Am J Health Syst Pharm. 2026. PMID: 41810511. doi:10.1093/ajhp/zxag076
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。