美国医疗系统静脉智能泵输注与报警评估
约30%的输注涉及至少1次报警,总体平均每次输注1.06次报警,有报警的输注平均每次3.51次报警,前4类报警占全部报警的81.8%。
要点
- 约30%的输注发生至少1次报警,有报警输注平均每次3.51次报警。
- 前4类报警占全部报警的81.8%,是报警管理优先关注对象。
- 抗肿瘤药、血制品、局麻药、免疫调节剂及PCA输注报警比例高于平均水平。
核心结论
- 研究描述了美国107家医院2年内静脉智能泵输注与报警的流行特征和类型分布。
- 报警频率与输注特征相关,可为优化报警管理、减轻报警疲劳提供依据。
研究设计与样本
- 回顾性描述性分析,基于大型数据集,覆盖美国107家医院2年内的静脉智能泵输注与报警数据。
主要发现
- 约30%的输注涉及1次或以上报警。
- 总体平均每次输注1.06次报警;有至少1次报警的输注平均每次3.51次报警。
- 前4类报警占全部报警的81.8%。
- 抗肿瘤药、血制品、局麻药、免疫调节剂及PCA输注的报警输注比例高于平均水平;单克隆抗体「bolus air in line」报警数显著高于平均水平;输注持续时间越长报警数越多。
临床怎么用
- 可依据前4类报警及高报警风险药物(抗肿瘤药、血制品、局麻药、免疫调节剂、单克隆抗体、PCA输注)优先开展报警优化。
- 对持续时间较长的输注应关注报警负担,评估减少不必要报警的措施。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: Intravenous smart pumps (IVSPs) are designed to safely and accurately deliver IV medications to patients. IVSP programming alerts are relatively well understood; however, little is known about IVSP alarms, their contribution to alarm fatigue, and how to minimize unnecessary ones. The purpose of this study was to describe the volume, frequency, and types of IVSP infusions and alarms that occurred over a 2-year period in 107 US hospitals. METHODS: A retrospective, descriptive analysis of a large-scale dataset containing IVSP infusion and alarm data was conducted. Additional variables, such as patient population, medication name, medication class, infusion type, and infusion duration were included in the analysis. RESULTS: More than 62 million infusions and 65 million alarms were analyzed. Approximately 30% of all infusions involved 1 or more alarms. Overall, there were an average of 1.06 alarms per infusion; for infusions with at least 1 alarm, there were 3.51 alarms per infusion. The top 4 alarm types accounted for 81.8% of all alarms. Antineoplastic agents, blood derivatives, local anesthetics, and immunomodulatory agents each had a higher-than-average percentage of infusions with at least 1 alarm, as did patient-controlled analgesia (PCA)-administered infusions. Monoclonal antibodies had a much higher-than-average number of "bolus air in line" alarms, specifically. A higher number of alarms was also observed for infusions of longer duration. CONCLUSION: This study provides valuable insights into the prevalence and characteristics of IVSP infusions and alarms. Key patterns in alarm frequency and alarm type and association between alarms and key infusion characteristics are also highlighted. These findings can inform efforts to optimize IVSP alarm management and decrease their contribution to alarm fatigue.
引用
- Armistead LT, Giuliano KK, Degnan DD, et al. Evaluating IV smart pump infusions and alarms in health systems across the United States. Am J Health Syst Pharm. 2026. PMID: 40971950. doi:10.1093/ajhp/zxaf233
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。