乳腺病理日常实践中全切片图像质量陷阱评估
本研究评估40160例乳腺全切片图像(WSI)与对应玻璃切片相比的组织缺失情况。组织缺失频率为2%至19%,缺失面积1%至70%,多数小于10%且位于边缘,均为脂肪组织。质量控制措施使WSI失败率降低七倍,无缺失组织导致最终诊断改变。
要点
- 乳腺WSI组织缺失频率2%至19%,缺失面积1%至70%。
- 多数缺失面积小于10%且位于周边,均为脂肪组织。
- 质量控制措施使WSI失败率降低七倍。
核心结论
- 乳腺WSI存在组织缺失但频率可变且诊断影响小。
- 需平衡WSI质量与扫描时间和文件大小。
- 实验室应自行评估风险并采取适当缓解措施。
研究设计与样本
- 检查40160例乳腺WSI并与对应玻璃切片比较。
- 评估组织缺失的存在、频率、位置、组织类型及临床影响。
- 考虑扫描时间、标本类型、WSI实施时间和质量控制措施。
主要发现
- 组织缺失频率范围为2%至19%。
- 缺失组织面积范围为1%至70%。
- 在大多数病例(>75%)中,缺失组织面积<10%且位于周边。
- 所有病例中缺失组织均为脂肪,伴或不伴少量包埋的正常乳腺实质。
- 空芯针活检标本WSI未发现组织缺失。
临床怎么用
- 乳腺WSI组织缺失频率可变但诊断影响小。
- 应平衡WSI质量与扫描时间和文件大小。
- 病理实验室应自行评估风险并提供适当缓解措施。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Using digitalized whole slide images (WSI) in routine histopathology practice is a revolutionary technology. This study aims to assess the clinical impacts of WSI quality and representation of the corresponding glass slides. 40,160 breast WSIs were examined and compared with their corresponding glass slides. The presence, frequency, location, tissue type, and the clinical impacts of missing tissue were assessed. Scanning time, type of the specimens, time to WSIs implementation, and quality control (QC) measures were also considered. The frequency of missing tissue ranged from 2% to 19%. The area size of the missed tissue ranged from 1-70%. In most cases (>75%), the missing tissue area size was <10% and peripherally located. In all cases the missed tissue was fat with or without small entrapped normal breast parenchyma. No missing tissue was identified in WSIs of the core biopsy specimens. QC measures improved images quality and reduced WSI failure rates by seven-fold. A negative linear correlation between the frequency of missing tissue and both the scanning time and the image file size was observed (p < 0.05). None of the WSI with missing tissues resulted in a change in the final diagnosis. Missing tissue on breast WSI is observed but with variable frequency and little diagnostic consequence. Balancing between WSI quality and scanning time/image file size should be considered and pathology laboratories should undertake their own assessments of risk and provide the relevant mitigations with the appropriate level of caution.
引用
- Atallah NM, Toss MS, Verrill C, et al. Potential quality pitfalls of digitalized whole slide image of breast pathology in routine practice. Mod Pathol. 2021. PMID: 34961765. doi:10.1038/s41379-021-01000-8
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。