全髋关节置换术中高孔隙度髋臼组件无菌性松动翻修分析

Revision for Aseptic Loosening of Highly Porous Acetabular Components in Primary Total Hip Arthroplasty: An Analysis of 20,993 Total Hip Replacements.
PubMed 文献 · J Arthroplasty · 2021-10-23

结果显示高孔隙度钛髋臼组件长期随访无菌性松动翻修率低,但Exeter V40/Tritanium组合相对POLARSTEM/R3(StikTite)的无菌性松动翻修风险存在显著差异,其余组合间无差异且无晚期无菌性松动翻修。

要点

  • 高孔隙度钛髋臼组件长期随访无菌性松动翻修率低。
  • Exeter V40/Tritanium相对POLARSTEM/R3(StikTite)的无菌性松动翻修风险存在显著差异。
  • 其他高孔隙度髋臼组件组合间无菌性松动翻修风险无差异,无晚期翻修。

核心结论

  • 高孔隙度涂层钛髋臼组件在长期随访中无菌性松动翻修率低。
  • 不同高孔隙度髋臼组件组合之间可能存在差异。

研究设计与样本

  • 纳入20,993例因骨关节炎行初次全髋关节置换术并使用高孔隙度髋臼组件的病例,共6种组合,均搭配高交联聚乙烯和32mm或36mm股骨头。
  • 主要结局为髋臼组件因无菌性松动而翻修,结果按患者年龄和性别调整。
  • 证据等级为Level III。

主要发现

  • 其他任何高孔隙度髋臼组件组合之间无差异。
  • 无晚期无菌性松动翻修。

关键数值与推荐

  • 主要结局指标为髋臼组件因无菌性松动而翻修。

临床怎么用

  • 高孔隙度钛髋臼组件长期无菌性松动翻修率低,可支持其临床应用。
  • 选择具体高孔隙度髋臼组件组合时需考虑潜在的无菌性松动翻修风险差异。

局限性

  • 本研究为登记系统回顾性分析,证据等级为Level III。
  • 结果仅针对因骨关节炎行初次全髋关节置换术并使用特定高孔隙度髋臼组件组合的病例。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Highly porous-coated titanium acetabular components have a high coefficient of friction and ultraporous surfaces to enhance bone ingrowth and osseointegration in total hip arthroplasty (THA). There have been concerns with the development of early radiolucent lines and aseptic loosening of highly porous acetabular components. It is unclear whether these concerns relate to a specific implant or the entire class. The aim of this study is to compare the revision rates for aseptic loosening of highly porous acetabular combinations in primary THA using data from a large joint replacement registry. METHODS: Data were retrieved from the Australian Orthopedic Association National Joint Replacement Registry for the study period September 1999 to December 2019. All primary THA procedures recorded and performed for osteoarthritis using the most common combinations for each highly porous acetabular component with highly cross-linked polyethylene and a 32-mm or 36-mm femoral head were included. The primary outcome measure was revision for aseptic loosening of the acetabular component. Results were adjusted for patient age and gender. RESULTS: There were 20,993 primary THA procedures performed for osteoarthritis using a highly porous acetabular component across 6 combinations. Relative to the POLARSTEM/R3 (StikTite), the Exeter V40/Tritanium had a significantly higher risk of revision for aseptic loosening of the acetabular component (hazard ratio 0.21, 95% confidence interval 0.06-0.74, P = .014). There was no difference between any other highly porous acetabular component combination and no late revisions for aseptic loosening. CONCLUSION: Highly porous-coated titanium acetabular components have low rates of aseptic loosening with long-term follow-up. A difference between components may exist. LEVEL OF EVIDENCE: Level III.