文献速递关节外科

短柄双锥度压配股骨柄517例初次全髋置换中期结果:无菌性松动率高

Mid-Term Results After 517 Primary Total Hip Arthroplasties With a Shortened and Shoulderless Double-Taper Press-Fit Stem: High Rates of Aseptic Loosening.
PubMed 文献 · J Arthroplasty · 2021-09-17

这项前瞻性单中心研究纳入486例患者共517例初次全髋关节置换术,使用MonoconMIS短柄双锥度压配股骨柄,平均随访5.29年。5年假体总生存率为95.2%,股骨柄生存率96.1%,翻修最常见原因为假体周围骨折(1.5%)、无菌性股骨柄松动(1.4%)和感染(1.0%)。无菌性松动仅见于Dorr A型股骨近端形态患者,作者不建议在该类患者中使用此柄。

要点

  • MonoconMIS短柄5年总生存率95.2%,股骨柄生存率96.1%,翻修率高于其他假体报道。
  • 无菌性股骨柄松动发生率1.4%,且仅发生于Dorr A型股骨近端形态患者。
  • WOMAC评分从术前49.57显著改善至5年时9.84,提示疗效充分。

核心结论

  • MonoconMIS股骨柄的翻修率高于其他假体报道的水平,其数据不支持在Dorr A型股骨近端形态患者中用于初次全髋关节置换。

研究设计与样本

  • 前瞻性单中心研究,纳入486例连续患者共517例初次全髋关节置换术,使用MonoconMIS短柄双锥度压配股骨柄,平均随访5.29年(SD 1.47)。

主要发现

  • 5年假体总生存率为95.2%,股骨柄生存率96.1%,髋臼杯生存率99.4%,孤立可动部件更换生存率99.0%。
  • 翻修最常见原因为假体周围骨折(1.5%)、无菌性股骨柄松动(1.4%)和感染(1.0%)。
  • WOMAC评分从基线49.57(SD 21.42)改善至术后1年13.33(SD 16.47)和术后5年9.84(SD 14.45)。
  • 无菌性股骨柄松动仅发生于Dorr A型股骨近端形态患者。

临床怎么用

  • 使用MonoconMIS短柄行初次全髋关节置换时,应警惕较高的翻修率,尤其对于Dorr A型股骨近端形态患者应避免使用。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Shorter double-taper stems with reduced lateral shoulders facilitate implantation via the muscle-sparing direct anterior approach and are becoming increasingly popular. We observed an unusually high number of cases of aseptic loosening with the use of a modified stem. Therefore, the aim of this prospective single-center study was to assess safety and efficacy of this cementless stem. METHODS: A total of 486 consecutive patients receiving 517 primary total hip arthroplasties using the MonoconMIS stem were prospectively followed up for a mean period of 5.29 years (standard deviation [SD], 1.47). Surgical and clinical data, complications, and revision surgeries were analyzed. The Western Ontario and McMaster Universities Arthritis Index (WOMAC) score was recorded before surgery and at one year and five years after surgery. RESULTS: The overall 5-year implant survival rate was 95.2%. The individual component survival rates were 96.1% for the stem, 99.4% for the acetabular cup, and 99.0% for the isolated mobile component exchange. The most common reasons for revision were periprosthetic fracture (1.5%), aseptic stem loosening (1.4%), and infection (1.0%). The WOMAC score improved significantly from 49.57 (SD, 21.42) at baseline to 13.33 (SD, 16.47) at one year and 9.84 (SD, 14.45) at five years after surgery. Aseptic stem loosening occurred only in patients with Dorr type A proximal femur morphology. CONCLUSION: The evaluated femoral stem is associated with revision rates higher than what has been reported for other implants. The WOMAC scores suggest adequate efficacy. Our data do not support the use of the MonoconMIS for primary total hip arthroplasty in patients with Dorr type A proximal femur morphology.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Wagner M, Schönthaler H, Endstrasser F, et al. Mid-Term Results After 517 Primary Total Hip Arthroplasties With a Shortened and Shoulderless Double-Taper Press-Fit Stem: High Rates of Aseptic Loosening. J Arthroplasty. 2021. PMID: 34625308. doi:10.1016/j.arth.2021.09.004

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。