仿生人工寰齿关节用于寰枢椎非融合:技术革新与初步生物力学研究

Atlantoaxial Non-Fusion Using Biomimetic Artificial Atlanto-Odontoid Joint: Technical Innovation and Initial Biomechanical Study.
PubMed 文献 · Spine (Phila Pa 1976) · 2021-10-06

本研究为体外生物力学研究,在10具新鲜人尸体颅颈标本上评估自研仿生人工寰齿关节(BAAOJ)置换对寰枢关节功能与稳定性的影响。结果显示BAAOJ植入后各方向活动度与中性区较完整状态轻度减小但无统计学差异(P>0.05),且较刚性钢板固定显著增加各方向活动度,尤其轴向旋转(P<0.05)。

要点

  • BAAOJ植入后寰枢椎各方向活动度与中性区较完整状态轻度减小,但差异无统计学意义(P>0.05)。
  • 与刚性钢板固定相比,BAAOJ植入显著增加各方向活动度与中性区,尤其轴向旋转(P<0.05)。
  • BAAOJ作为非融合装置可保留类似正常状态的屈伸、侧屈和轴向旋转活动度,并稳定寰枢复合体。

核心结论

  • BAAOJ置换可保留寰枢椎屈伸、侧屈和轴向旋转活动度,且能稳定寰枢复合体,BAAOJ本身具有良好的初始稳定性。

研究设计与样本

  • 体外生物力学研究,对10具新鲜人尸体颅颈标本依次进行完整状态、BAAOJ关节成形术后、BAAOJ疲劳测试后、齿突切除术后及前路刚性钢板固定后测试。

主要发现

  • BAAOJ植入与完整状态相比,各方向活动度和中性区轻度减小(P>0.05)。
  • 与刚性钢板固定相比,BAAOJ植入显著增加各方向活动度和中性区,尤其轴向旋转(P<0.05)。

临床怎么用

  • BAAOJ作为非融合装置,有望在治疗寰枢椎不稳相关疾病时保留寰枢椎活动度并维持稳定。

局限性

  • 本研究为体外生物力学研究,证据等级为4级,样本量为10具尸体标本。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

STUDY DESIGN: A biomechanical in vitro investigation. OBJECTIVE: To evaluate the function and stability of self-designed biomimetic artificial atlanto-odontoid joint (BAAOJ) replacement on the atlantoaxial joint. SUMMARY OF BACKGROUND DATA: Upper cervical fusion surgery is a common treatment for various atlantoaxial disorders, and favorable clinical outcome has been achieved. However, the fusion surgery results in loss of atlantoaxial motion as well as adjacent segments degeneration, reducing the quality of life of patients and might produce severe neurological symptoms. Non-fusion technology is expected to solve the above problems, but various designed devices have certain defects and are still in the exploratory phase. MATERIALS AND METHODS: Biomechanical tests were conducted on 10 fresh human cadaveric craniocervical specimens in the following sequence: 1) intact condition, 2) after the BAAOJ arthroplasty, 3) after BAAOJ fatigue test, 4) after odontoidect-omy, and 5) after anterior rigid plate fixation. Three-dimensional movements of the C1-C2 segment were evaluated to investigate the function and stability of BAAOJ arthroplasty compared with the intact condition after the BAAOJ fatigue test, odontoidect-omy, and rigid plate fixation. RESULTS: Comparing the BAAOJ implantation to the intact state, the range of motion and neutral zone were slightly reduced in all directions (P > 0.05). Compared with the rigid plate fixation, the BAAOJ implantation significantly increased the range of motion and neutral zone in all directions, especially in the axial rotation (P < 0.05). CONCLUSION: We designed a BAAOJ for correcting atlantoaxial disorders arising from atlantoaxial instability. As a non-fusion device, the most critical feature of BAAOJ replacement is the retention of flexion-extension, lateral bending, and axial rotation range of motion similar to the normal state. It can also stabilize the atlantoaxial complex, and the BAAOJ itself has a good initial stability.Level of Evidence: 4.

来源:Spine (Phila Pa 1976),查看原文。版权归原作者所有。