文献速递脊柱外科

桁架式椎间融合器骨矿化与脊柱融合评估:绵羊有限元分析及体内验证

Bone Mineralization and Spinal Fusion Evaluation of a Truss-based Interbody Fusion Device: Ovine Finite Element Analysis with Confirmatory In Vivo Outcomes.
PubMed 文献 · Spine (Phila Pa 1976) · 2022-01-01

本研究通过有限元分析(FEA)模拟生理负荷下不同支柱直径(0.75、1.0、1.25、1.5 mm)桁架式椎间融合器的应变分布,并在绵羊腰椎椎间融合模型中验证0.75 mm与1.5 mm支柱的融合结局。FEA显示支柱直径与峰值应变幅度呈负相关,体内实验证实0.75 mm支柱融合评分显著更高、融合率更高且矿化骨组织更多。

要点

  • FEA显示支柱直径越小,峰值应变幅度越大,1.0、1.25、1.5 mm支柱峰值应变分别较0.75 mm低36%、60%、73%。
  • 体内实验0.75 mm支柱融合评分在3个月时显著高于1.5 mm支柱,并在后续各时间点保持显著优势(P < 0.001)。
  • 0.75 mm支柱融合率在3、6、12个月分别为95%、100%、100%,1.5 mm支柱分别为72.7%、86.4%、95.8%。

核心结论

  • 较小直径支柱的融合器表现出更快的融合巩固和更积极的骨整合,验证了FEA的预测。

研究设计与样本

  • 研究设计为有限元分析与绵羊体内脊柱椎间融合研究。
  • FEA模拟生理负荷条件,评估支柱直径0.75、1.0、1.25、1.5 mm的应变分布。
  • 体内研究比较0.75 mm与1.5 mm支柱直径融合器的融合评分和骨组织形态计量学指标,随访3、6、12个月。

主要发现

  • FEA显示支柱直径与峰值应变幅度呈负相关,1.0、1.25、1.5 mm支柱峰值应变较0.75 mm分别低36%、60%、73%。
  • 体内实验0.75 mm支柱融合评分在3个月时显著高于1.5 mm支柱,并在后续各时间点保持显著更高(P < 0.001 for all comparisons)。
  • 融合率:0.75 mm支柱在3、6、12个月分别为95%、100%、100%;1.5 mm支柱分别为72.7%、86.4%、95.8%。
  • 各随访时间点,较细支柱的矿化骨组织更多,纤维/软骨组织更少。

临床怎么用

  • 在桁架式椎间融合器设计中,较小直径支柱可能通过增大应变幅度促进更快的融合和骨整合。

局限性

  • 证据等级为4级。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

STUDY DESIGN: Finite element analysis (FEA) and in vivo ovine spinal interbody fusion study. OBJECTIVE: To determine comparative load-induced strain amplitudes, bone mineralization and fusion outcomes associated with different diameter struts in a truss-based interbody fusion device. SUMMARY OF BACKGROUND DATA: Additive manufacturing technology has been employed to develop implants that actively participate in the fusion process. The truss device enables the optimal transfer of compressive and tensile stresses via the struts. Mechanobiologic principles postulate that strut diameter can be regulated to allow different magnitudes of strain distribution within the struts which may affect fusion rates. METHODS: Modeling of strain distributions as a function of strut diameter (0.75, 1.0, 1.25, and 1.5 mm) employed FEA that simulated physiologic loading conditions. A confirmatory in vivo ovine lumbar spinal interbody fusion study compared fusion scores and bone histomorphometric variables for cages with 0.75 and 1.5 mm strut diameters. Outcomes were compared at 3-, 6-, and 12-month follow-up intervals. RESULTS: FEA showed an inverse association between strut diameter and peak strain amplitude. Cages with 1.0, 1.25, and 1.5 mm struts had peak strain values that were 36%, 60%, and 73% lower than the 0.75 mm strut strain value. In vivo results showed the mean fusion score for the 0.75 mm diameter strut cage was significantly greater by 3-months versus the 1.5 mm strut cage, and remained significantly higher at each subsequent interval (P < 0.001 for all comparisons). Fusion rates were 95%, 100%, and 100% (0.75 mm) and 72.7%, 86.4%, and 95.8% (1.5 mm) at 3, 6, and 12 months. Thinner struts had greater mineralized bone tissue and less fibrous/chondral tissue than the thicker struts at each follow-up. CONCLUSION: Validating FEA estimates, cages with smaller diameter struts exhibited more rapid fusion consolidation and more aggressive osseointegration compared with cages with larger diameters struts.Level of Evidence: 4.

来源:Spine (Phila Pa 1976),查看原文。版权归原作者所有。

引用

  1. Kiapour A, Seim HB, Atkinson BL, et al. Bone Mineralization and Spinal Fusion Evaluation of a Truss-based Interbody Fusion Device: Ovine Finite Element Analysis with Confirmatory In Vivo Outcomes. Spine (Phila Pa 1976). 2022. PMID: 34593736. doi:10.1097/BRS.0000000000004256

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。