中央锁定型胫骨托与骨水泥分离与聚乙烯垫片背面变形相关

Tibial tray debonding from the cement mantle is associated with deformation of the backside of polyethylene tibial inserts.
PubMed 文献 · Bone Joint J · 2021-09-03

本研究对114例取出的固定平台全膝关节置换(TKA)组件进行分析,测量聚乙烯(PE)垫片的磨损与变形及胫骨托背面骨水泥覆盖百分比。结果发现,中央锁定型(Attune、NexGen)胫骨托的骨水泥覆盖显著少于外周锁定型(Triathlon、PFC),且其PE垫片平面度中位数显著更大,并呈现特征性背面变形模式。

要点

  • 中央锁定型胫骨托的骨水泥覆盖显著低于外周锁定型(p=0.001)。
  • 中央锁定型PE垫片平面度中位数显著大于外周锁定型(205 vs 85微米;p<0.001)。
  • 中央锁定型PE垫片出现特征性背面变形,外侧边缘下移,边缘成为主要关节接触区。

核心结论

  • 中央锁定机制的取出TKA胫骨托更易与骨水泥层分离。
  • 这些设计的PE垫片呈现特征性变形模式,可能与制造工艺相关并在体内加重。
  • 该变形模式与关节边缘的PE磨损相关,可能增加界面摩擦扭矩。

研究设计与样本

  • 回顾性取出物分析,来自单中心取出中心,共检查114例固定平台TKA组件。
  • 其中76例使用当代PE垫片,15例使用老一代PE垫片进行对比。
  • 使用坐标测量机和光学显微镜测量PE磨损与变形,量化胫骨托背面骨水泥覆盖百分比。
  • 采用回归模型识别与骨水泥覆盖量相关的变量。

主要发现

  • Attune和NexGen(中央锁定)胫骨托的骨水泥覆盖显著少于Triathlon和PFC(外周锁定)胫骨托(p=0.001)。
  • 中央锁定型PE垫片平面度中位数为205微米,显著大于外周锁定型的85微米(p<0.001)。
  • Attune和NexGen垫片具有特征性背面变形模式,PE外侧边缘下移,PE边缘成为主要关节接触区。

临床怎么用

  • 对于使用中央锁定型胫骨托的TKA,应关注骨水泥覆盖是否充分及PE垫片背面变形迹象。
  • 随访中若发现PE边缘磨损或胫骨托松动,需考虑中央锁定机制相关的分离风险。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: The aim of this study was to investigate whether wear and backside deformation of polyethylene (PE) tibial inserts may influence the cement cover of tibial trays of explanted total knee arthroplasties (TKAs). METHODS: At our retrieval centre, we measured changes in the wear and deformation of PE inserts using coordinate measuring machines and light microscopy. The amount of cement cover on the backside of tibial trays was quantified as a percentage of the total surface. The study involved data from the explanted fixed-bearing components of four widely used contemporary designs of TKA (Attune, NexGen, Press Fit Condylar (PFC), and Triathlon), revised for any indication, and we compared them with components that used previous generations of PE. Regression modelling was used to identify variables related to the amount of cement cover on the retrieved trays. RESULTS: A total of 114 explanted fixed-bearing TKAs were examined. This included 76 used with contemporary PE inserts which were compared with 15 used with older generation PEs. The Attune and NexGen (central locking) trays were found to have significantly less cement cover than Triathlon and PFC trays (peripheral locking group) (p = 0.001). The median planicity values of the PE inserts used with central locking trays were significantly greater than of those with peripheral locking inserts (205 vs 85 microns; p < 0.001). Attune and NexGen inserts had a characteristic pattern of backside deformation, with the outer edges of the PE deviating inferiorly, leaving the PE margins as the primary areas of articulation. CONCLUSION: Explanted TKAs with central locking mechanisms were significantly more likely to debond from the cement mantle. The PE inserts of these designs showed characteristic patterns of deformation, which appeared to relate to the manufacturing process and may be exacerbated in vivo. This pattern of deformation was associated with PE wear occurring at the outer edges of the articulation, potentially increasing the frictional torque generated at this interface. Cite this article: Bone Joint J 2021;103-B(12):1791-1801.

来源:Bone Joint J,查看原文。版权归原作者所有。

引用

  1. Bhalekar RM, Nargol ME, Shyam N, et al. Tibial tray debonding from the cement mantle is associated with deformation of the backside of polyethylene tibial inserts. Bone Joint J. 2021. PMID: 34474593. doi:10.1302/0301-620X.103B.BJJ-2020-2555.R2

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。