文献速递关节外科

单一股骨柄设计机械辅助缝隙腐蚀翻修术后的结局

Outcomes Following Revision for Mechanically Assisted Crevice Corrosion in a Single Femoral Design.
PubMed 文献 · J Arthroplasty · 2021-08-18

回顾性队列研究纳入621例金属对聚乙烯全髋关节置换,因机械辅助缝隙腐蚀翻修率为11.6%,平均翻修时间6.6年。翻修术后平均3.2年主要并发症率11.6%、再手术率5.8%,最终随访髋关节功能障碍与骨关节炎结局评分关节置换版平均83.2, hip主观值平均77.6,均较术前显著改善(P < .001)。

要点

  • 该植入物组合的机械辅助缝隙腐蚀翻修率为11.6%,平均翻修时间6.6年。
  • 翻修术后平均3.2年主要并发症率11.6%,再手术率5.8%。
  • 翻修术后髋关节功能评分显著改善,但并发症率仍较高。

核心结论

  • 金属对聚乙烯全髋关节置换中机械辅助缝隙腐蚀的发生率可能高于既往报道,尤其在特定植入物组合下。
  • 因机械辅助缝隙腐蚀行翻修术可获得良好结局,但高临床怀疑、早期发现和早期翻修可能是成功的关键。

研究设计与样本

  • 回顾性队列研究,纳入552例连续患者共621例金属对聚乙烯初次全髋关节置换,所有患者均使用同一植入物组合(Accolade I股骨柄/钴铬低摩擦离子处理股骨头)。
  • 识别其中因机械辅助缝隙腐蚀诊断行翻修术的患者,收集人口学、手术和实验室数据,随访自翻修手术起计算,末次随访评估髋关节功能障碍与骨关节炎结局评分关节置换版和髋主观值。

主要发现

  • 机械辅助缝隙腐蚀翻修率为11.6%,平均翻修时间6.6(±2.4)年。
  • 翻修患者(n = 69)术前血清钴铬比平均3.5(±2.4),其中8例发生严重柄锥失效。
  • 翻修术后平均3.2(±1.9)年总体主要并发症率为11.6%,再手术率为5.8%。
  • 末次随访时髋关节功能障碍与骨关节炎结局评分关节置换版平均83.2(±15.6),髋主观值平均77.6(±17.4),两者较术前均显著增加(P < .001)。

临床怎么用

  • 对使用特定植入物组合的金属对聚乙烯全髋关节置换患者,应保持对机械辅助缝隙腐蚀的高临床怀疑。
  • 早期发现并尽早行翻修术可能有助于获得良好结局。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Mechanically assisted crevice corrosion (MACC) is a described complication following metal-on-polyethylene (MoP) total hip arthroplasty (THA). The literature regarding outcomes following revision for MACC suggests that complication rates are high. The purpose of this investigation is to add to this literature with the largest reported series to date. METHODS: This is a retrospective cohort study of 552 consecutive patients who underwent 621 MoP primary THAs. We identified patients who subsequently underwent revision THA for a diagnosis of MACC. All patients were implanted with the same implant combination (Accolade I stem/cobalt-chromium low friction ion treatment femoral head). Patient demographic, surgical, and laboratory data were collected. Follow-up was calculated from the revision surgery and Hip Disability and Osteoarthritis Outcome Score Joint Replacement and hip subjective values (HSV) were examined at final follow-up. Descriptive statistics were performed. RESULTS: The revision rate for MACC was 11.6% and mean time to revision was 6.6 (±2.4) years. Revised patients (n = 69) had a mean preoperative serum cobalt-chromium ratio of 3.5 (±2.4). There were 8 cases of gross trunnion failure. At mean 3.2 (±1.9) years following revision, the overall major complication rate was 11.6% with a 5.8% reoperation rate. At final follow-up, mean Hip Disability and Osteoarthritis Outcome Score Joint Replacement scores were 83.2 (±15.6) and mean hip subjective value was 77.6 (±17.4). Revision resulted in significant increases in both parameters (P < .001). CONCLUSION: The incidence of MACC in MoP THA is likely higher than previously reported, particularly for certain implant combinations. Revision surgery for MACC can achieve good outcomes but a high clinical suspicion with early detection and revision is likely key to success.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Wilson JM, Broggi MS, Oladeji P, et al. Outcomes Following Revision for Mechanically Assisted Crevice Corrosion in a Single Femoral Design. J Arthroplasty. 2021. PMID: 34481694. doi:10.1016/j.arth.2021.08.010