脊柱畸形手术中影响椎弓根螺钉置入扭矩的因素

Factors Affecting Pedicle Screw Insertional Torque in Spine Deformity Surgery.
PubMed 文献 · Spine (Phila Pa 1976) · 2021-01-01

该回顾性观察研究比较了成人脊柱畸形(ASD)与青少年特发性脊柱侧凸(AIS)患者L1-L3椎弓根螺钉置入扭矩,发现椎弓根与螺钉直径均与扭矩正相关,ASD获得刚性固定的最佳螺钉/椎弓根比值接近但小于1,而AIS约为1至1.25。

要点

  • 椎弓根直径和螺钉直径均与置入扭矩正相关。
  • ASD最佳螺钉/椎弓根比值接近但小于1。
  • AIS最佳螺钉/椎弓根比值约为1至1.25。

核心结论

  • 提出螺钉/椎弓根比值概念,ASD与AIS的最佳比值不同,有助于术前选择合适螺钉直径。

研究设计与样本

  • 回顾性观察研究,连续患者,比较ASD患者68例324枚螺钉与AIS患者32例58枚螺钉的L1-L3置入扭矩。

主要发现

  • ASD和AIS患者中椎弓根直径与螺钉直径均与置入扭矩呈正相关。
  • ASD获得刚性固定的最佳螺钉/椎弓根比值接近但小于1。
  • AIS获得刚性固定的最佳螺钉/椎弓根比值约为1至1.25。

关键数值与推荐

  • ASD最佳螺钉/椎弓根比值接近但小于1。
  • AIS最佳螺钉/椎弓根比值约为1至1.25。

临床怎么用

  • 螺钉/椎弓根比值可用于术前规划,根据畸形类型选择合适直径螺钉以获得刚性固定。

局限性

  • 证据等级为4级。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

STUDY DESIGN: Retrospective observational study of consecutive patients. OBJECTIVE: We sought to: (1) clarify the key factors predominantly associated with the insertional torque of pedicle screws; (2) compare the optimal factors for pedicle screw insertion to obtain rigid screw fixation in patients with adult spinal deformity (ASD) and in those with adolescent idiopathic scoliosis (AIS); (3) determine the optimal screw/pedicle ratio (S/P) to obtain rigid pedicle screw fixation. SUMMARY OF BACKGROUND DATA: Rigid pedicle screw fixation is mandatory to perform corrections for spinal deformities properly and to allow successful fusion after surgery. The fixation depends mainly on screw position accuracy and patient bone quality. Traditionally, spinal surgeons have decided the screw size, trajectory, and tapping size based on their intuition. Insertional torque has been indicated as useful to predict screw fixation strength, and is correlated with screw pullout strength and frequency of postoperative screw loosening. METHODS: We compared insertion torque at L1-L3 levels of 324 screws in 68 patients with ASD and 58 screws in 32 patients with AIS. We assessed the association between screw/pedicle ratio and insertion torque by constructing a spline curve. RESULTS: Pedicle and screw diameter correlated positively with insertion torque in patients with either ASD or AIS. The optimal screw/pedicle ratio to obtain rigid pedicle screw fixation in patients with ASD was close to, but less than one, and, by contrast, was about 1 to 1.25 in patients with AIS. CONCLUSION: We propose the concept of an optimal S/P ratio for obtaining rigid pedicle screw fixation during spinal corrective surgery, which is different for patients with ASD and patients with AIS. The S/P ratio is useful for deciding the appropriate diameter screw for each case in preoperative planning.Level of Evidence: 4.

来源:Spine (Phila Pa 1976),查看原文。版权归原作者所有。