最大尺寸金属头是否增加维生素E交联聚乙烯磨损?随机对照试验两年结果

Does the use of the largest possible metal head increase the wear of vitamin E-doped cross-linked polyethylene? Two-year results from a randomized controlled trial.
PubMed 文献 · Bone Joint J · 2021-01-01

单盲随机对照试验纳入96例全髋关节置换患者,比较最大可能钴铬头(36-44mm)与标准32mm头对维生素E交联聚乙烯内衬磨损的影响。两年RSA显示两组近端头穿透量无显著差异(p=0.323),体积磨损率亦无差异(p=0.848),大尺寸头未增加线性磨损,但体积磨损有偏高趋势,需长期评估。

要点

  • 最大尺寸金属头与32mm头在两年时近端头穿透量无显著差异(p=0.323)。
  • 两组稳态体积磨损率无显著差异(6.1 vs 3.5 mm³/年,p=0.848)。
  • 大尺寸头体积磨损有偏高趋势,需长期随访才能定论。

核心结论

  • 与32mm头相比,使用最大可能金属头在两年时未增加维生素E交联聚乙烯磨损。
  • 两组线性磨损可忽略,体积磨损率均很低。
  • 大尺寸头体积磨损有偏高趋势,需长期评估才能确定头尺寸与磨损的关联。

研究设计与样本

  • 单盲随机对照试验,96例患者(43例女性),中位年龄63岁(IQR 57-69)。
  • 随机分配至最大可能模块化股骨头(36-44mm)配最薄内衬或标准32mm头。
  • 所有患者均使用维生素E交联聚乙烯内衬和钴铬头。
  • 主要结局为两年时放射立体照相分析(RSA)测量的近端头穿透量。
  • 次要结局包括体积磨损、髋臼周围透亮线和患者报告结局。

主要发现

  • 两年时每组44例患者可进行RSA评估。
  • 最大可能头组中位两年总近端头穿透量为-0.02mm(IQR -0.09至0.07;p=0.548)。
  • 32mm头组中位两年总近端头穿透量为-0.01mm(IQR -0.07至0.10;p=0.525)。
  • 两组差异无统计学意义(p=0.323),均未显示磨合期。
  • 中位稳态体积磨损率分别为6.1mm³/年(IQR -59至57)和3.5mm³/年(-21至34),组间无差异(p=0.848)。

关键数值与推荐

  • 主要结局:两年近端头穿透量,最大头组-0.02mm vs 32mm组-0.01mm,p=0.323。
  • 稳态体积磨损率:最大头组6.1mm³/年 vs 32mm组3.5mm³/年,p=0.848。

临床怎么用

  • 在维生素E交联聚乙烯内衬中,使用最大可能金属头(36-44mm)两年内未增加线性或体积磨损。
  • 大尺寸头体积磨损有偏高趋势,临床选择头尺寸时需结合长期随访证据。

局限性

  • 随访仅两年,无法确定长期磨损差异。
  • 大尺寸头体积磨损有偏高趋势,需更长期评估。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: We aimed to investigate if the use of the largest possible cobalt-chromium head articulating with polyethylene acetabular inserts would increase the in vivo wear rate in total hip arthroplasty. METHODS: In a single-blinded randomized controlled trial, 96 patients (43 females), at a median age of 63 years (interquartile range (IQR) 57 to 69), were allocated to receive either the largest possible modular femoral head (36 mm to 44 mm) in the thinnest possible insert or a standard 32 mm head. All patients received a vitamin E-doped cross-linked polyethylene insert and a cobalt-chromium head. The primary outcome was proximal head penetration measured with radiostereometric analysis (RSA) at two years. Secondary outcomes were volumetric wear, periacetabular radiolucencies, and patient-reported outcomes. RESULTS: At two years, 44 patients in each group were available for RSA assessment. The median total two-year proximal head penetration was -0.02 mm (IQR -0.09 to 0.07; p = 0.548) for the largest possible head and -0.01 mm (IQR -0.07 to 0.10; p = 0.525) for 32 mm heads. Their difference was not statistically significant (p = 0.323). Neither group demonstrated a bedding-in period. The median steady-state volumetric wear rates were 6.1 mm3/year (IQR -59 to 57) and 3.5 mm3/year (-21 to 34) respectively, and did not differ between the groups (p = 0.848). There were no statistically significant differences in periacetabular radiolucencies or patient-reported outcomes. CONCLUSION: The use of the largest possible metal head did not increase vitamin E-doped cross-linked polyethylene wear compared with 32 mm heads at two years. Linear wear was negligible and volumetric wear rates were very low in both head size groups. There was a tendency towards higher values of volumetric wear in large heads that warrants longer-term evaluation before any definite conclusions about the association between head size and wear can be drawn. Cite this article: Bone Joint J 2021;103-B(7):1206-1214.