骨水泥内骨水泥股骨假体翻修:两种锥形滑配设计的中期随访比较

Cement-in-cement femoral component revision : a comparison of two different taper-slip designs with medium-term follow up.
PubMed 文献 · Bone Joint J · 2021-01-01

本研究比较Exeter短翻修柄与C-Stem AMT高偏距1号假体在骨水泥内骨水泥股骨翻修中的临床与影像学结果。97例患者(Exeter 50例,C-Stem AMT 47例)平均随访9.7年,两组OHS、WOMAC、SF-12均显著改善,Exeter组腿长显著较短(平均-4 mm vs 0 mm),共16例(15%)再次翻修股骨假体,无因无菌性松动翻修,Exeter组有2例假体骨折。结论认为两种双/三锥形滑配设计在正确指征下安全可靠。

要点

  • 骨水泥内骨水泥翻修使用Exeter或C-Stem AMT假体中期临床评分均显著改善。
  • Exeter组腿长显著较短(平均-4 mm),C-Stem AMT组为0 mm。
  • 共16例(15%)再次翻修股骨假体,无因无菌性松动翻修,Exeter组2例假体骨折。

核心结论

  • 使用Exeter或C-Stem AMT假体行骨水泥内骨水泥翻修的中期结果良好。
  • 双或三锥形滑配设计的骨水泥内骨水泥翻修在正确指征下是安全可靠的技术。

研究设计与样本

  • 回顾性比较研究,基于前瞻性收集的数据库,纳入2005年1月至2013年1月因各种指征行骨水泥内骨水泥股骨翻修且随访至少2年的患者。
  • 所有手术由资深作者完成,使用Exeter短翻修柄或C-Stem AMT高偏距1号假体,每年进行临床和影像学评估。
  • 共97例患者符合纳入标准(Exeter 50例,C-Stem AMT 47例),两组人口学数据无显著差异,平均随访9.7年。

主要发现

  • 两组Oxford髋关节评分(OHS)、WOMAC和SF-12评分均显著改善。
  • Exeter组腿长显著较短,平均-4 mm,而C-Stem AMT组为0 mm。
  • Exeter组1例有早期影像学松动证据。
  • 共16例患者(15%)再次翻修股骨假体(C-Stem AMT组7例,Exeter组9例),无因无菌性松动翻修。
  • Exeter组有2例股骨假体骨折。

关键数值与推荐

  • 纳入标准:2005年1月至2013年1月行骨水泥内骨水泥翻修且随访至少2年的患者。
  • Exeter组腿长平均-4 mm,C-Stem AMT组0 mm,差异显著。
  • 再次翻修股骨假体共16例(15%),其中C-Stem AMT组7例,Exeter组9例。
  • Exeter组股骨假体骨折2例。

临床怎么用

  • 对于正确指征的股骨假体翻修,骨水泥内骨水泥技术使用Exeter或C-Stem AMT假体可获得良好的中期临床评分改善。
  • 需注意Exeter组腿长平均缩短4 mm及假体骨折风险,术中应加以关注。

局限性

  • 本研究为单术者回顾性比较,样本量97例,随访9.7年,可能存在选择偏倚。
  • 原文未报告两组翻修指征的具体分布及假体骨折的具体处理细节。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: Cement-in-cement revision of the femoral component represents a widely practised technique for a variety of indications in revision total hip arthroplasty. In this study, we compare the clinical and radiological outcomes of two polished tapered femoral components. METHODS: From our prospectively collated database, we identified all patients undergoing cement-in-cement revision from January 2005 to January 2013 who had a minimum of two years' follow-up. All cases were performed by the senior author using either an Exeter short revision stem or the C-Stem AMT high offset No. 1 prosthesis. Patients were followed-up annually with clinical and radiological assessment. RESULTS: A total of 97 patients matched the inclusion criteria (50 Exeter and 47 C-Stem AMT components). There were no significant differences between the patient demographic data in either group. Mean follow-up was 9.7 years. A significant improvement in Oxford Hip Score (OHS), Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), and 12-item Short-Form Survey (SF-12) scores was observed in both cohorts. Leg lengths were significantly shorter in the Exeter group, with a mean of -4 mm in this cohort compared with 0 mm in the C-Stem AMT group. One patient in the Exeter group had early evidence of radiological loosening. In total, 16 patients (15%) underwent further revision of the femoral component (seven in the C-Stem AMT group and nine in the Exeter group). No femoral components were revised for aseptic loosening. There were two cases of femoral component fracture in the Exeter group. CONCLUSION: Our series shows promising mid-term outcomes for the cement-in-cement revision technique using either the Exeter or C-Stem AMT components. These results demonstrate that cement-in-cement revision using a double or triple taper-slip design is a safe and reliable technique when used for the correct indications. Cite this article: Bone Joint J 2021;103-B(7):1215-1221.

来源:Bone Joint J,查看原文。版权归原作者所有。

引用

  1. Kennedy IW, Ng NYB, Young D, et al. Cement-in-cement femoral component revision : a comparison of two different taper-slip designs with medium-term follow up. Bone Joint J. 2021. PMID: 34192939. doi:10.1302/0301-620X.103B7.BJJ-2020-1953.R2

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。