模块化双动髋臼杯金属内衬的微动与腐蚀:取回分析

Fretting and corrosion of metal liners from modular dual mobility constructs : a retrieval analysis.
PubMed 文献 · Bone Joint J · 2021-01-01

该取回研究对60例模块化双动(MDM)髋臼杯金属内衬进行微动和腐蚀评分,并分析3例最严重腐蚀植入物的腐蚀产物及48例假体周围组织学。结果显示88%内衬存在微动、97%存在腐蚀,6例组织学见巨噬细胞和颗粒提示对腐蚀碎屑的炎症反应,均与高腐蚀评分相关。

要点

  • MDM金属内衬-外壳界面普遍存在微动(88%)和腐蚀(97%)。
  • 微动最常见于锥部唇缘与穹顶过渡区。
  • 高腐蚀评分病例的假体周围组织可见对腐蚀碎屑的炎症反应。

核心结论

  • MDM内衬与外壳界面发生腐蚀,且可伴随局部组织反应,长期使用中该问题可能具有临床重要性。

研究设计与样本

  • 取回研究:对60例取回的MDM金属内衬进行微动和腐蚀的主观评分。
  • 对3例最严重腐蚀植入物取表面腐蚀产物行扫描电镜成像和傅里叶变换红外光谱分析。
  • 对48例假体周围组织进行组织学检查。

主要发现

  • 88%(53/60)的取回内衬存在微动,97%(58/60)存在腐蚀。
  • 微动最常位于锥部唇缘与穹顶区域的过渡处。
  • 48例组织学检查中6例见巨噬细胞和颗粒,反映对腐蚀碎屑的先天炎症反应,均与高腐蚀评分取回组件相关。

临床怎么用

  • 临床随访MDM髋臼杯时应关注金属内衬-外壳界面的腐蚀及局部组织反应风险。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: Dual mobility implants in total hip arthroplasty are designed to increase the functional head size, thus decreasing the potential for dislocation. Modular dual mobility (MDM) implants incorporate a metal liner (e.g. cobalt-chromium alloy) in a metal shell (e.g. titanium alloy), raising concern for mechanically assisted crevice corrosion at the modular liner-shell connection. We sought to examine fretting and corrosion on MDM liners, to analyze the corrosion products, and to examine histologically the periprosthetic tissues. METHODS: A total of 60 retrieved liners were subjectively scored for fretting and corrosion. The corrosion products from the three most severely corroded implants were removed from the implant surface, imaged using scanning electron microscopy, and analyzed using Fourier-transform infrared spectroscopy. RESULTS: Fretting was present on 88% (53/60) of the retrieved liners, and corrosion was present on 97% (58/60). Fretting was most often found on the lip of the taper at the transition between the lip and the dome regions. Macrophages and particles reflecting an innate inflammatory reaction to corrosion debris were noted in six of the 48 cases for which periprosthetic tissues were examined, and all were associated with retrieved components that had high corrosion scores. CONCLUSION: Our results show that corrosion occurs at the interface between MDM liners and shells and that it can be associated with reactions in the local tissues, suggesting continued concern that this problem may become clinically important with longer-term use of these implants. Cite this article: Bone Joint J 2021;103-B(7):1238-1246.

来源:Bone Joint J,查看原文。版权归原作者所有。

引用

  1. Hemmerling KJ, Weitzler L, Bauer TW, et al. Fretting and corrosion of metal liners from modular dual mobility constructs : a retrieval analysis. Bone Joint J. 2021. PMID: 34192940. doi:10.1302/0301-620X.103B7.BJJ-2020-0221.R1

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。