文献速递脊柱外科骨质疏松与代谢性骨病

双能X线吸收法不能反映骨质疏松患者的骨结构:腰椎DXA与椎体活检的对比

Dual-energy X-ray Absorptiometry Does Not Represent Bone Structure in Patients with Osteoporosis: A Comparison of Lumbar Dual-Energy X-Ray Absorptiometry with Vertebral Biopsies.
PubMed 文献 · Spine (Phila Pa 1976) · 2021-01-01

这项前瞻性横断面探索性研究对比了66例骨质疏松患者(33例椎体骨折组,33例对照组)的腰椎DXA与椎体活检micro-CT分析。骨折组DXA和micro-CT参数均差于对照组,但骨折组DXA与micro-CT参数无相关性;对照组仅部分参数相关。结论提示腰椎DXA不能定性或定量反映微结构骨架构,不是评估脊柱骨质量的可靠工具。

要点

  • 骨折组DXA与micro-CT参数无相关性,DXA不能反映骨微结构。
  • 对照组仅骨体积分数与aBMD(r=0.51)、骨小梁数与aBMD(r=0.56)相关。
  • 腰椎DXA不能可靠评估脊柱骨质量。

核心结论

  • 腰椎DXA在定性或定量上均不能代表微结构骨架构,不是评估脊柱骨质量的可靠工具。

研究设计与样本

  • 前瞻性横断面探索性研究,纳入66例患者,骨折组与对照组各33例。
  • 术前所有患者接受腰椎DXA扫描(L1-L4)。
  • 术中取非骨折椎体(优先L3)活检,用micro-CT分析骨量与骨质量。

主要发现

  • 骨折组术前DXA结果差于对照组:面积骨密度(aBMD)0.95 vs. 1.31,T值-1.97 vs. 0.92,均P < 0.001。
  • micro-CT证实两组间定量参数有差异:骨/总体积0.09 vs. 0.12,P < 0.001。
  • micro-CT证实两组间定性参数有差异:连接指数15.73 vs. 26.67,P < 0.001;结构模型指数2.66 vs. 2.27,P < 0.001;骨小梁数2.11 vs. 2.28,P = 0.014。
  • 骨折组DXA结果与micro-CT参数无相关性。
  • 对照组部分参数有相关性:骨/总体积与aBMD r = 0.51,P = 0.005;骨小梁数与aBMD r = 0.56,P = 0.001。

临床怎么用

  • 临床不应仅凭腰椎DXA评估骨质疏松患者的脊柱骨质量。
  • 对椎体骨折风险高的患者,需结合其他手段评估骨微结构。

局限性

  • 证据等级为3级。
  • 探索性研究,样本量66例。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

STUDY DESIGN: Prospective cross-sectional exploratory study. OBJECTIVE: To evaluate the correlation between in vivo lumbar dual-energy x-ray absorptiometry (DXA) and parameters of bone architecture in micro-computed tomography (micro-CT) in patients with osteoporosis. SUMMARY OF BACKGROUND DATA: DXA is the current diagnostic standard for evaluating osteoporosis. However, there are various concerns regarding its validity, especially in the spine. No study has so far investigated whether in vivo DXA correlates with the actual lumbar bone architecture. METHODS: Lumbar DXA scans were compared with micro-CT analysis of vertebral biopsies in patients with osteoporotic vertebral fractures (fracture group) and those without (control group). Preoperatively, all patients underwent a DXA scan (L1-L4). Intraoperative biopsies from nonfractured vertebrae (preferably L3) were analyzed by micro-CT regarding bone quantity and quality. The groups were compared regarding differences in DXA and micro-CT results. In each group, a correlation analysis was performed between DXA and micro-CT. RESULTS: The study included 66 patients (33 per group). Preoperative DXA results were worse in the fracture group than the control group (areal bone mineral density [aBMD] 0.95 vs. 1.31, T-score -1.97 vs. 0.92, each P < 0.001). Micro-CT analysis confirmed differences regarding quantitative parameters (bone/total volume: 0.09 vs. 0.12, P < 0.001) and qualitative parameters (connectivity index: 15.73 vs. 26.67, P < 0.001; structure model index: 2.66 vs. 2.27, P < 0.001; trabecular number: 2.11 vs. 2.28, P = 0.014) of bone architecture between both groups. The DXA results did not correlate with micro-CT parameters in the fracture group. In the control group, correlations were found for some parameters (bone/total volume vs. aBMD: r = 0.51, P = 0.005; trabecular number vs. aBMD: r = 0.56, P = 0.001). CONCLUSION: These data constitute the first comparison of DXA measurements with microstructural analysis of vertebral biopsies in patients with osteoporosis. Our results indicate that lumbar DXA neither qualitatively nor quantitatively represents microstructural bone architecture and is therefore not a reliable tool for the evaluation of bone quality in the spine.Level of Evidence: 3.

来源:Spine (Phila Pa 1976),查看原文。版权归原作者所有。

引用

  1. Pumberger M, Palmowski Y, Strube P, et al. Dual-energy X-ray Absorptiometry Does Not Represent Bone Structure in Patients with Osteoporosis: A Comparison of Lumbar Dual-Energy X-Ray Absorptiometry with Vertebral Biopsies. Spine (Phila Pa 1976). 2021. PMID: 34100839. doi:10.1097/BRS.0000000000003917

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。