钴水平与假瘤特征因金属离子来源而异:模块化股骨颈与金属对金属关节的对比
结果显示假瘤特征在CoCrMN与MoM组间不同,正常血清钴水平不能排除假瘤存在,建议对该类CoCrMN假体随访时常规行MARS MRI。
要点
- CoCrMN组假瘤患者中血清钴升高者占57.9%,MoM组占73.7%。
- CoCrMN组假瘤多位于髋关节外侧,MoM组多位于前内侧。
核心结论
- CoCrMN设计与MoM关节患者的假瘤特征不同。
- 正常血清钴水平不能排除假瘤存在。
- 建议对CoCrMN假体患者随访时常规行MARS MRI。
研究设计与样本
- 回顾性评估2005至2013年由一位外科医生植入的190例初次全髋关节置换术。
- 63例使用钛颈组件和MoM关节,10例使用钛颈组件和聚乙烯 liner。
- 所有患者记录血清离子水平,假瘤分级和位置由肌肉骨骼放射科医生测量。
主要发现
- 钴水平与索引关节置换术后时间不相关。
- CoCrMN组假瘤患者中钴水平升高者11/19(57.9%),MoM组为14/19(73.7%)。
- CoCrMN组假瘤更常位于髋关节外侧,MoM关节置换后多位于前内侧。
关键数值与推荐
- CoCrMN组假瘤患者钴水平升高比例57.9%(11/19)。
- MoM组假瘤患者钴水平升高比例73.7%(14/19)。
临床怎么用
- 对使用CoCrMN假体的患者,即使血清钴水平正常也应考虑行MARS MRI以排除假瘤。
- 随访中需关注假瘤的可能,尤其位于髋关节外侧的假瘤。
- 常规MARS MRI应纳入该类假体患者的随访方案。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: This study examines cobalt levels and pseudotumor characteristics in a non-recalled modular femoral prosthesis. We sought to determine relationships between serum cobalt levels and pseudotumors following modular and metal-on-metal (MoM) total hip arthroplasty (THA), the location and size of pseudotumors, and changes to pseudotumor grade over time. METHODS: We retrospectively evaluated a cohort of 190 primary THAs implanted with a dual-modular femoral stem by one surgeon from 2005 to 2013. One hundred seventeen THAs used a cobalt-chromium modular neck (CoCrMN) component and non-MoM articulation, 63 THAs had a titanium neck component and MoM articulation, and 10 had a titanium neck component and polyethylene liner. Serum ion levels were recorded for all patients. Pseudotumor grade and location were measured by musculoskeletal-trained radiologists. RESULTS: Cobalt levels did not correlate to time after index arthroplasty. In the CoCrMN group, cobalt levels were elevated in 11 of 19 (57.9%) patients with pseudotumor compared to 14 of 19 (73.7%) in the MoM group. Pseudotumors were more often located lateral to the hip joint in the CoCrMN group compared to anteromedial following MoM arthroplasty. In CoCrMN THAs, pseudotumors were visualized in 12 of 51 (23.5%) initial MARS MRIs within 40 months compared to 5 of 66 (7.6%) performed later (P = .02). CONCLUSION: Pseudotumor characteristics differ between patients with CoCrMN designs vs those with an MoM articulation. Normal serum cobalt levels did not exclude the presence of a pseudotumor and routine MARS MRI should be included in follow-up of patients with this CoCrMN prostheses.
引用
- McConnell ZA, Stambough JB, Barnes CL, et al. Cobalt Levels and Pseudotumor Characteristics Vary due to Metal Ion Source: Modular Femoral Neck vs Metal-on-Metal Articulations. J Arthroplasty. 2021. PMID: 34172347. doi:10.1016/j.arth.2021.05.018
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。