现代骨水泥固定平台后稳定型膝关节假体翻修率意外偏高
单中心回顾性研究比较166例ATTUNE骨水泥固定平台后稳定型全膝关节置换与511例其他设计假体,ATTUNE组翻修率11.5%(19例),翻修待定7%(12例),胫骨组件松动占翻修的90%;参考队列翻修率2.6%、翻修待定1.5%,差异显著。
要点
- ATTUNE组翻修率11.5%,参考队列仅2.6%,差异显著(p<0.001)。
- 翻修主要原因为胫骨组件松动,占ATTUNE组翻修的90%。
- 骨水泥类型和术者与翻修无显著关联(p=0.988和p=0.550)。
核心结论
- 该新型全膝关节置换设计相比参考队列翻修率意外偏高,胫骨组件脱粘是主要失败原因。
- 需要对这一特定组件及植入技术进行更长期随访研究。
研究设计与样本
- 单医院回顾性研究,纳入166例(154例患者)连续ATTUNE骨水泥固定平台后稳定型全膝关节置换,由5位术者完成。
- 参考队列为同期同院同术者完成的511膝(470例患者)其他7个厂商设计的全膝关节置换。
- 两组年龄、性别、BMI和随访时间无显著差异。
主要发现
- ATTUNE组19膝(11.5%)已翻修,平均翻修时间30.3个月(SD 15),其中17膝(90%)见胫骨组件松动。
- ATTUNE组另有12膝(7%)翻修待定。
- 参考队列翻修13膝(2.6%)、翻修待定8膝(1.5%),均显著低于ATTUNE组(p<0.001)。
- 参考队列仅2膝因胫骨组件松动翻修。
- 翻修或翻修待定的31膝与其余135膝在患者特征、骨水泥类型(p=0.988)和术者(p=0.550)上无差异。
临床怎么用
- 使用该特定版本ATTUNE胫骨组件时需警惕早期胫骨组件松动导致的翻修风险。
- 研究中植入的ATTUNE胫骨组件版本已由制造商进行修改。
局限性
- 本研究为单中心回顾性设计。
- 原文未报告其他局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIMS: To establish our early clinical results of a new total knee arthroplasty (TKA) tibial component introduced in 2013 and compare it to other designs in use at our hospital during the same period. METHODS: This is a retrospective study of 166 (154 patients) consecutive cemented, fixed bearing, posterior-stabilized (PS) TKAs (ATTUNE) at one hospital performed by five surgeons. These were compared with a reference cohort of 511 knees (470 patients) of other designs (seven manufacturers) performed at the same hospital by the same surgeons. There were no significant differences in age, sex, BMI, or follow-up times between the two cohorts. The primary outcome was revision performed or pending. RESULTS: In total, 19 (11.5%) ATTUNE study TKAs have been revised at a mean 30.3 months (SD 15), and loosening of the tibial component was seen in 17 of these (90%). Revision is pending in 12 (7%) knees. There was no difference between the 31 knees revised or with revision pending and the remaining 135 study knees in terms of patient characteristics, type of bone cement (p = 0.988), or individual surgeon (p = 0.550). In the reference cohort, there were significantly fewer knees revised (n = 13, 2.6%) and with revision pending (n = 8, 1.5%) (both p < 0.001), and only two had loosening of the tibial component as the reason for revision. CONCLUSION: This new TKA design had an unexpectedly high early rate of revision compared with our reference cohort of TKAs. Debonding of the tibial component was the most common reason for failure. Additional longer-term follow-up studies of this specific component and techniques for implantation are warranted. The version of the ATTUNE tibial component implanted in this study has undergone modifications by the manufacturer. Cite this article: Bone Joint J 2021;103-B(6 Supple A):137-144.
引用
- Lachiewicz PF, Steele JR, Wellman SS, et al. Unexpected high rate of revision of a modern cemented fixed bearing modular posterior-stabilized knee arthroplasty. Bone Joint J. 2021. PMID: 34053294. doi:10.1302/0301-620X.103B6.BJJ-2020-1956.R1