文献速递关节外科

基于三维超声的膝关节快速三维骨骼重建新方法

A novel approach to 3D bone creation in minutes : 3D ultrasound.
PubMed 文献 · Bone Joint J · 2021-01-01

本研究评估超声作为无辐射成像模态重建膝关节三维解剖用于术前模板规划。通过电磁追踪与射频信号处理提取骨边界,结合统计形状模型重建患者特异性表面。在17具尸体上对比CT模型,股骨和胫骨RMSE分别为1.07 mm和1.02 mm,标定和假体尺寸选择具较高准确性,操作者经验无显著影响。

要点

  • 三维超声骨模型与CT模型高度匹配,股骨和胫骨RMSE约1 mm。
  • 股骨标定准确,胫骨标定误差略高;假体尺寸选择多数正确或仅差一号。
  • 操作者经验(中等与熟练)对输出无差异,具临床快速规划潜力。

核心结论

  • 三维超声骨模型与CT模型高度匹配,适用于术前规划,且无辐射。

研究设计与样本

  • 采用带电磁追踪器的超声系统采集股骨和胫骨射频信号,实时提取骨边界并融合追踪信息生成三维点云,通过统计形状模型重建患者特异性表面。
  • 在17具尸体上比较三维超声模型与CT模型的准确性,并比较15名操作者的扫描以评估操作者变异影响。

主要发现

  • 股骨RMSE为1.07 mm(SD 0.15),胫骨RMSE为1.02 mm(SD 0.13)。
  • 股骨标定:经上髁轴1.07°(SD 0.65°),后髁轴0.73°(SD 0.41°),远端髁轴0.96°(SD 0.89°),内侧AP 1.22 mm(SD 0.69),外侧AP 1.21 mm(SD 1.02)。
  • 胫骨标定误差略高:后倾轴1.92°(SD 1.31°),结节轴1.91°(SD 1.24°)。
  • 假体尺寸选择:90%股骨和60%胫骨尺寸正确,其余仅差一个尺寸。
  • 中等与熟练操作者之间无差异。

临床怎么用

  • 三维超声可无辐射快速可视化骨骼,用于临床就诊期间术前规划与假体尺寸选择。
  • 潜在扩展应用包括三维动态韧带平衡和术中配准用于机器人与导航系统。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: The objective of this study is to assess the use of ultrasound (US) as a radiation-free imaging modality to reconstruct 3D anatomy of the knee for use in preoperative templating in knee arthroplasty. METHODS: Using an US system, which is fitted with an electromagnetic (EM) tracker that is integrated into the US probe, allows 3D tracking of the probe, femur, and tibia. The raw US radiofrequency (RF) signals are acquired and, using real-time signal processing, bone boundaries are extracted. Bone boundaries and the tracking information are fused in a 3D point cloud for the femur and tibia. Using a statistical shaping model, the patient-specific surface is reconstructed by optimizing bone geometry to match the point clouds. An accuracy analysis was conducted for 17 cadavers by comparing the 3D US models with those created using CT. US scans from 15 users were compared in order to examine the effect of operator variability on the output. RESULTS: The results revealed that the US bone models were accurate compared with the CT models (root mean squared error (RM)S: femur, 1.07 mm (SD 0.15); tibia, 1.02 mm (SD 0.13). Additionally, femoral landmarking proved to be accurate (transepicondylar axis: 1.07° (SD 0.65°); posterior condylar axis: 0.73° (SD 0.41°); distal condylar axis: 0.96° (SD 0.89°); medial anteroposterior (AP): 1.22 mm (SD 0.69); lateral AP: 1.21 mm (SD 1.02)). Tibial landmarking errors were slightly higher (posterior slope axis: 1.92° (SD 1.31°); and tubercle axis: 1.91° (SD 1.24°)). For implant sizing, 90% of the femora and 60% of the tibiae were sized correctly, while the remainder were only one size different from the required implant size. No difference was observed between moderate and skilled users. CONCLUSION: The 3D US bone models were proven to be closely matched compared with CT and suitable for preoperative planning. The 3D US is radiation-free and offers numerous clinical opportunities for bone visualization rapidly during clinic visits, to enable preoperative planning with implant sizing. There is potential to extend its application to 3D dynamic ligament balancing, and intraoperative registration for use with robots and navigation systems. Cite this article: Bone Joint J 2021;103-B(6 Supple A):81-86.

来源:Bone Joint J,查看原文。版权归原作者所有。

引用

  1. Mahfouz MR, Abdel Fatah EE, Johnson JM, et al. A novel approach to 3D bone creation in minutes : 3D ultrasound. Bone Joint J. 2021. PMID: 34053279. doi:10.1302/0301-620X.103B6.BJJ-2020-2455.R1