全髋关节置换术中无影像计算机导航不同注册平面的准确性评估

Evaluating Alternate Registration Planes for Imageless, Computer-Assisted Navigation During Total Hip Arthroplasty.
PubMed 文献 · J Arthroplasty · 2021-06-02

一项前瞻性观察研究纳入94例初次全髋关节置换术患者,比较无影像导航下四种注册平面(冠状面、纵向面、腰骶面、髋肩面)与术后X线测量的髋臼杯位置准确性。结果显示髋肩面对前倾角测量最准确,腰骶面对外展角测量误差显著;结论为近似身体长轴的平面可提供准确参考,髋肩面提供最准确一致的测量。

核心结论

  • 使用近似身体长轴的任何平面进行患者注册均可准确测量术中髋臼杯位置,其中髋肩面提供最准确且一致的前倾角测量。

研究设计与样本

  • 前瞻性观察研究,纳入94例接受初次全髋关节置换术并使用无影像导航辅助的患者,比较四种注册平面与术后X线测量。

关键数值与推荐

  • 注册平面包括:患者冠状面(标准方法)、手术台长轴(纵向面)、腰骶脊柱(腰骶面)、经大转子和肩胛盂窝的平面(髋肩面)。

临床怎么用

  • 在侧卧位全髋关节置换术中,无影像导航注册可优先选择髋肩面以提高前倾角测量准确性,避免使用腰骶面导致外展角高估。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Imageless computer navigation improves component placement accuracy in total hip arthroplasty (THA), but variations in the registration process are known to impact final accuracy measurements. We sought to evaluate the registration accuracy of an imageless navigation device during THA performed in the lateral decubitus position. METHODS: A prospective, observational study of 94 patients undergoing a primary THA with imageless navigation assistance was conducted. Patient position was registered using 4 planes of reference: the patient's coronal plane (standard method), the long axis of the surgical table (longitudinal plane), the lumbosacral spine (lumbosacral plane), and the plane intersecting the greater trochanter and glenoid fossa (hip-shoulder plane). Navigation measurements of cup position for each plane were compared to measurements from postoperative radiographs. RESULTS: Mean inclination from radiographs (41.5° ± 5.6°) did not differ significantly from inclination using the coronal plane (40.9° ± 3.9°, P = .39), the hip-shoulder plane (42.4° ± 4.7°, P = .26), or the longitudinal plane (41.2° ± 4.3°, P = .66). Inclination measured using the lumbosacral plane (45.8° ± 4.3°) differed significantly from radiographic measurements (P < .0001). Anteversion measured from radiographs (mean: 26.1° ± 5.4°) did not differ significantly from the hip-shoulder plane (26.6° ± 5.2°, P = .50). All other planes differed significantly from radiographs: coronal (22.6° ± 6.8°, P = .001), lumbosacral (32.5° ± 6.4°, P < .0001), and longitudinal (23.7° ± 5.2°, P < .0001). CONCLUSION: Patient registration using any plane approximating the long axis of the body provided a frame of reference that accurately measured intraoperative cup position. Registration using a plane approximating the hip-shoulder axis, however, provided the most accurate and consistent measurement of acetabular component position.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Vigdorchik JM, Sculco PK, Inglis AE, et al. Evaluating Alternate Registration Planes for Imageless, Computer-Assisted Navigation During Total Hip Arthroplasty. J Arthroplasty. 2021. PMID: 34154856. doi:10.1016/j.arth.2021.05.037

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。