机器人辅助全膝关节置换术相比传统技术提高组件定位准确性和精确性
本研究回顾性比较了同一术者连续220例初次全膝关节置换术,其中103例机械工具TKA(M-TKA)和96例机器人辅助TKA(RA-TKA)。RA-TKA在股骨、胫骨组件定位、胫骨后倾和下肢机械轴对线方面偏离计划均显著小于M-TKA(均p<0.001),且聚乙烯厚度偏差更小(1.4 mm vs 2.7 mm,p<0.001)。结论认为RA-TKA在组件定位和聚乙烯厚度规划上更准确和精确。
要点
- RA-TKA组股骨、胫骨定位及下肢机械轴偏离计划均显著小于M-TKA组(均p<0.001)。
- RA-TKA组聚乙烯厚度偏差显著更小(1.4 mm vs 2.7 mm,p<0.001)。
- RA-TKA组需股骨远端再截骨的比例显著更低(10% vs 22%,p=0.033)。
核心结论
- RA-TKA在组件定位和最终聚乙烯厚度规划方面显著更准确和精确。
- 未来需研究这种提高的准确性和精确性是否影响临床结局。
研究设计与样本
- 计划冠状面组件对线和整体下肢对线均为0°机械轴,计划胫骨后倾M-TKA为2°、RA-TKA为1.5°,计划聚乙烯厚度为9 mm。
主要发现
- RA-TKA组股骨定位平均偏差0.9°(SD 1.2°),M-TKA组1.7°(SD 1.1°),差异显著(p<0.001)。
- RA-TKA组胫骨定位平均偏差0.3°(SD 0.9°),M-TKA组1.3°(SD 1.0°),差异显著(p<0.001)。
- RA-TKA组胫骨后倾平均偏差-0.3°(SD 1.3°),M-TKA组1.7°(SD 1.1°),差异显著(p<0.001)。
- RA-TKA组下肢机械轴平均偏差1.0°(SD 1.7°),M-TKA组2.7°(SD 1.9°),差异显著(p<0.001)。
- RA-TKA组聚乙烯厚度平均偏差1.4 mm(SD 1.6),M-TKA组2.7 mm(SD 2.2),差异显著(p<0.001)。
关键数值与推荐
- RA-TKA组需股骨远端再截骨10例(10%),M-TKA组22例(22%),差异显著(p=0.033)。
- 计划冠状面股骨和胫骨组件对线及整体下肢对线均为0±3°。
临床怎么用
- 对于追求组件定位准确性和可重复性的初次全膝关节置换术,RA-TKA可能有助于减少对线偏离和聚乙烯厚度偏差。
- RA-TKA可能降低股骨远端再截骨的需求。
局限性
- 本研究为单术者回顾性研究,样本来自连续病例,可能存在选择偏倚。
- 未评估准确性和精确性提高对临床结局的影响。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIMS: Robotic-assisted total knee arthroplasty (RA-TKA) is theoretically more accurate for component positioning than TKA performed with mechanical instruments (M-TKA). Furthermore, the ability to incorporate soft-tissue laxity data into the plan prior to bone resection should reduce variability between the planned polyethylene thickness and the final implanted polyethylene. The purpose of this study was to compare accuracy to plan for component positioning and precision, as demonstrated by deviation from plan for polyethylene insert thickness in measured-resection RA-TKA versus M-TKA. METHODS: A total of 220 consecutive primary TKAs between May 2016 and November 2018, performed by a single surgeon, were reviewed. Planned coronal plane component alignment and overall limb alignment were all 0° to the mechanical axis; tibial posterior slope was 2°; and polyethylene thickness was 9 mm. For RA-TKA, individual component position was adjusted to assist gap-balancing but planned coronal plane alignment for the femoral and tibial components and overall limb alignment remained 0 ± 3°; planned tibial posterior slope was 1.5°. Mean deviations from plan for each parameter were compared between groups for positioning and size and outliers were assessed. RESULTS: In all, 103 M-TKAs and 96 RA-TKAs were included. In RA-TKA versus M-TKA, respectively: mean femoral positioning (0.9° (SD 1.2°) vs 1.7° (SD 1.1°)), mean tibial positioning (0.3° (SD 0.9°) vs 1.3° (SD 1.0°)), mean posterior tibial slope (-0.3° (SD 1.3°) vs 1.7° (SD 1.1°)), and mean mechanical axis limb alignment (1.0° (SD 1.7°) vs 2.7° (SD 1.9°)) all deviated significantly less from the plan (all p < 0.001); significantly fewer knees required a distal femoral recut (10 (10%) vs 22 (22%), p = 0.033); and deviation from planned polyethylene thickness was significantly less (1.4 mm (SD 1.6) vs 2.7 mm (SD 2.2), p < 0.001). CONCLUSION: RA-TKA is significantly more accurate and precise in planning both component positioning and final polyethylene insert thickness. Future studies should investigate whether this increased accuracy and precision has an impact on clinical outcomes. The greater accuracy and reproducibility of RA-TKA may be important as precise new goals for component positioning are developed and can be further individualized to the patient. Cite this article: Bone Joint J 2021;103-B(6 Supple A):74-80.