单侧后路近顶椎撑开装置(ApiFix)非融合治疗青少年特发性脊柱侧凸的高失败率

High Failure Rates of a Unilateral Posterior Peri-Apical Distraction Device (ApiFix) for Fusionless Treatment of Adolescent Idiopathic Scoliosis.
PubMed 文献 · J Bone Joint Surg Am · 2021-01-01

这项前瞻性队列研究纳入20例青少年特发性脊柱侧凸(AIS)患者,采用ApiFix装置行非融合矫形。平均随访3.4年,主弯从术前45.4°降至术后2周31.4°、末次随访31.0°;

要点

  • ApiFix非融合装置初期矫形效果尚可,但随访期内未见持续撑开。
  • 6例翻修术中培养出Cutibacterium acnes。

核心结论

  • ApiFix单侧后路近顶椎自撑开棘轮棒初期临床与影像学结果有希望,但未见撑开效应,2年内严重不良事件发生率高,在该机构被认为不可接受,尽管已获FDA批准。

研究设计与样本

  • 前瞻性队列研究,纳入20例AIS患者(Risser 1-4级、Lenke 1或5型、主弯Cobb角40°至55°、Bunnell侧凸计旋转<15°),采用ApiFix装置治疗。
  • 患者平均年龄14.8±1.4岁,平均随访3.4±1.0年。
  • 证据等级:治疗性IV级。

主要发现

  • 主弯平均从术前45.4°降至术后2周31.4°、末次随访31.0°。
  • 次弯平均从术前31.3°降至术后2周26.1°、末次随访24.2°。
  • 翻修术中所有患者均见金属沉积,6例培养出Cutibacterium acnes。
  • 因高失败率研究提前终止。

关键数值与推荐

  • 纳入标准:Risser 1-4级、Lenke 1或5型、主弯Cobb角40°至55°、Bunnell侧凸计旋转<15°。
  • 严重并发症定义:需翻修手术的不良事件。

临床怎么用

  • 该装置初期矫形满意但缺乏持续撑开,且严重并发症发生率高,临床应用需谨慎评估。
  • 翻修术中应关注金属沉积及低毒性感染(如Cutibacterium acnes)的可能。

局限性

  • 样本量小(20例),随访时间平均3.4年。
  • 研究因高失败率提前终止,可能影响长期结果评估。
  • 证据等级为治疗性IV级。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Conventional surgical treatment for adolescent idiopathic scoliosis (AIS) consists of correction of the spinal deformity with rigid spinal instrumentation and fusion. Less-invasive and fusionless surgery could potentially improve patient outcomes. The purpose of the present study was to evaluate the efficacy of a recently U.S. Food and Drug Administration (FDA)-approved posterior peri-apical self-distracting device (ApiFix) that is designed to gradually correct the deformity without spinal fusion. METHODS: In a prospective cohort study of 20 patients with AIS (Risser stage 1-4; Lenke 1 or 5; major curve Cobb angle, 40° to 55°; and Bunnell scoliometer rotation, <15°) were managed with the ApiFix device. Clinical and radiographic performance was assessed. RESULTS: Twenty patients with a mean age (and standard deviation) of 14.8 ± 1.4 years were followed for a mean of 3.4 ± 1.0 years. The average major curve was reduced from 45.4° preoperatively to 31.4° at 2 weeks postoperatively and 31.0° at the time of the latest follow-up. The average minor curve measured 31.3° preoperatively, 26.1° at 2 weeks postoperatively, and 24.2° at the time of the latest follow-up. Ten patients had serious complications that required revision surgery, including osteolysis (n = 6), screw and/or rod breakage (n = 2), failure of the ratchet mechanism (n = 1), and pain without explainable cause (n = 1). During revision surgery, metallosis was observed in all patients and cultures showed growth of Cutibacterium acnes in 6 patients. Because of the high failure rate, the study was terminated early. CONCLUSIONS: The use of the unilateral peri-apical concave self-distracting ratchet rod initially was associated with promising clinical and radiographic results. However, no distraction was observed and the high rate of serious adverse events within 2 years was considered to be unacceptable for further clinical application of this device in our institution, despite recent FDA approval. LEVEL OF EVIDENCE: Therapeutic Level IV. See Instructions for Authors for a complete description of levels of evidence.