严重损伤CoCrMo股骨头损伤模式与微观结构和材料损失的关系
本研究对108例严重损伤的锻造CoCrMo全髋关节置换股骨头进行回收分析,发现化学主导的柱状损伤占48%,且柱状损伤者材料损失体积更大(中位数1.2 mm³ vs 0.6 mm³,p=0.03)。合金微观结构中的带状偏析与材料损失增加相关(中位数1.1 mm³ vs 0.2 mm³,p=0.02),提示制造过程中应避免带状偏析。
要点
- 严重损伤股骨头中化学主导柱状损伤常见,占48%。
- 柱状损伤者材料损失体积大于单纯机械损伤者。
- 合金带状偏析与材料损失增加相关,应避免。
核心结论
- 化学主导的柱状损伤是严重损伤股骨头中体积损失更大的明确指标。
- 体积材料损失强烈依赖于锻造CoCrMo合金中的带状偏析。
- 制造过程中应避免带状偏析以减少材料损失和腐蚀产物释放。
研究设计与样本
- 回顾性分析2004年6月至2019年6月取出的锻造CoCrMo模块化股骨头,共1002例,其中190例(19%)为严重损伤,排除后最终纳入108例。
- 使用改良Goldberg视觉评分系统评估损伤,用光学坐标测量机识别损伤模式和体积材料损失,并分析合金微观结构特征。
主要发现
- 化学驱动柱状损伤见于48%(52/108)股骨头,34%(37/108)为柱状损伤合并压印,12%(13/108)为柱状损伤合并均匀微动磨损,2%(2/108)为广泛柱状损伤无法确认潜在机械损伤模式。
- 柱状损伤者材料损失中位数(范围)为1.2 mm³(0.2-11.7),大于单纯机械损伤者的0.6 mm³(0-20.7),p=0.03。
- 所有股骨头材料损失中位数为0.9 mm³(0-20.7)。
- 合金带状偏析者材料损失中位数(范围)为1.1 mm³(0-20.7),均质 microstructure 者为0.2 mm³(0-4.1),p=0.02。
- 硬相含量和晶粒尺寸与材料损失无相关性。
关键数值与推荐
- 柱状损伤组材料损失中位数1.2 mm³(0.2-11.7),机械损伤组0.6 mm³(0-20.7),p=0.03。
- 带状偏析组材料损失中位数1.1 mm³(0-20.7),均质组0.2 mm³(0-4.1),p=0.02。
临床怎么用
- 对于使用锻造CoCrMo股骨头的全髋关节置换,应注意合金微观结构中的带状偏析,可能增加材料损失和不良组织反应风险。
- 制造过程中避免带状偏析可能减少植入物体内碎屑释放。
局限性
- 仅纳入严重损伤的回收股骨头,可能存在选择偏倚。
- 样本量有限(108例),且为回顾性研究。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Fretting and corrosion in metal-on-polyethylene total hip arthoplasty (THA) modular junctions can cause adverse tissue reactions that are responsible for 2% to 5% of revision surgeries. Damage within cobalt-chromium-molybdenum (CoCrMo) alloy femoral heads can progress chemically and mechanically, leading to damage modes such as column damage, imprinting, and uniform fretting damage. At present, it is unclear which of these damage modes are most detrimental and how they may be linked to implant alloy metallurgy. The alloy microstructure exhibits microstructural features such as grain boundaries, hard phases, and segregation bands, which may enable different damage modes, higher material loss, and the potential risk of adverse local tissue reactions. QUESTIONS/PURPOSES: In this study, we asked: (1) How prevalent is chemically dominated column damage compared with mechanically dominated damage modes in severely damaged metal-on-polyethylene THA femoral heads made from wrought CoCrMo alloy? (2) Is material loss greater in femoral heads that underwent column damage? (3) Do material loss and the presence of column damage depend on alloy microstructure as characterized by grain size, hard phase content, and/or banding? METHODS: Surgically retrieved wrought CoCrMo modular femoral heads removed between June 2004 and June 2019 were scored using a modified version of the Goldberg visually based scoring system. Of the total 1002 heads retrieved over this period, 19% (190 of 1002) were identified as severely damaged, exhibiting large areas of fretting scars, black debris, pits, and/or etch marks. Of these, 43% (81 of 190) were excluded for metal-on-metal articulations, alternate designs (such as bipolar, dual-mobility, hemiarthroplasty, metal adaptor sleeves), or previous sectioning of the implant for past studies. One sample was excluded retroactively as metallurgical analysis revealed that it was made of cast alloy, yielding a total of 108 for further analysis. Information on patient age (57 ± 11 years) and sex (56% [61 of 108] were males), reason for removal, implant time in situ (99 ± 78 months), implant manufacturer, head size, and the CoCrMo or titanium-based stem alloy pairing were collected. Damage modes and volumetric material loss within the head tapers were identified using an optical coordinate measuring machine. Samples were categorized by damage mode groups by column damage, imprinting, a combination of column damage and imprinting, or uniform fretting. Metallurgical samples were processed to identify microstructural characteristics of grain size, hard phase content, and banding. Nonparametric Mann-Whitney U and Kruskal-Wallis statistical tests were used to examine volumetric material loss compared with damage mode and microstructural features, and linear regression was performed to correlate patient- and manufacturer-specific factors with volumetric material loss. RESULTS: Chemically driven column damage was seen in 48% (52 of 108) of femoral heads, with 34% (37 of 108) exhibiting a combination of column damage and imprinting, 12% (13 of 108) of heads displaying column damage and uniform fretting, and 2% (2 of 108) exhibiting such widespread column damage that potentially underlying mechanical damage modes could not be verified. Implants with column damage showed greater material loss than those with mechanically driven damage alone, with median (range) values of 1.2 mm3 (0.2 to 11.7) versus 0.6 mm3 (0 to 20.7; p = 0.03). Median (range) volume loss across all femoral heads was 0.9 mm3 (0 to 20.7). Time in situ, contact area, patient age, sex, head size, manufacturer, and stem alloy type were not associated with volumetric material loss. Banding of the alloy microstructure, with a median (range) material loss of 1.1 mm3 (0 to 20.7), was associated with five times higher material loss compared with those with a homogeneous microstructure, which had a volume loss of 0.2 mm3 (0 to 4.1; p = 0.02). Hard phase content and grain size showed no correlation with material loss. CONCLUSION: Chemically dominated column damage was a clear indicator of greater volume loss in this study sample of 108 severely damaged heads. Volumetric material loss strongly depended on banding (microstructural segregations) within the alloy. Banding of the wrought CoCrMo microstructure should be avoided during the manufacturing process to reduce volumetric material loss and the release of corrosion products to the periprosthetic tissue. CLINICAL RELEVANCE: Approximately 30% of THAs rely on wrought CoCrMo femoral heads. Most femoral heads in this study exhibited a banded microstructure that was associated with larger material loss and the occurrence of chemically dominated column damage. This study suggests that elimination of banding from the alloy could substantially reduce the release of implant debris in vivo, which could potentially also reduce the risk of adverse local tissue reactions to implant debris.