对称与非对称胫骨假体全膝关节置换术的两年RSA迁移结果:一项随机对照试验

Two-year RSA migration results of symmetrical and asymmetrical tibial components in total knee arthroplasty: a randomized controlled trial.
PubMed 文献 · Bone Joint J · 2021-01-01

本研究为随机对照RSA试验,纳入75例骨水泥型后稳定型全膝关节置换术,比较非对称胫骨设计(Persona PS)与对称胫骨设计(NexGen LPS)的两年迁移模式和临床结局。线性混合效应模型分析31例非对称和38例对称TKA的RSA数据,未发现胫骨或股骨假体平均MTPM迁移模式差异;两年时非对称设计胫骨假体平均MTPM为0.93 mm,对称设计为1.00 mm;股骨假体分别为1.04 mm和1.14 mm。其他迁移参数及临床和PROM测量均无显著差异。结论认为两种设计的假体迁移相当,长期无菌性松动风险可能相当。

要点

  • 非对称与对称胫骨假体TKA的两年RSA迁移模式无差异。
  • 两年胫骨假体MTPM:非对称0.93 mm,对称1.00 mm。
  • 临床结局和患者报告结局测量无显著差异。

核心结论

  • 非对称胫骨假体设计与经验证的对称胫骨假体设计具有相当的假体迁移,提示两种设计的长期无菌性松动风险相当。

研究设计与样本

  • 随机对照放射立体分析(RSA)试验,纳入75例骨水泥型后稳定型全膝关节置换术。
  • 使用线性混合效应模型分析31例非对称设计和38例对称设计TKA的RSA数据。
  • 在术后一周及术后3、6、12、24个月评估假体迁移、临床结局和患者报告结局测量。

主要发现

  • 线性混合效应模型未显示非对称与对称设计在胫骨或股骨假体平均MTPM迁移模式上的差异。
  • 术后两年非对称设计胫骨假体平均MTPM为0.93 mm,对称设计为1.00 mm。
  • 术后两年非对称设计股骨假体平均MTPM为1.04 mm,对称设计为1.14 mm。
  • 其他迁移参数、临床测量和PROM测量均未观察到显著差异。

关键数值与推荐

  • 术后两年胫骨假体平均MTPM:非对称设计0.93 mm,对称设计1.00 mm。
  • 术后两年股骨假体平均MTPM:非对称设计1.04 mm,对称设计1.14 mm。

临床怎么用

  • 非对称胫骨假体设计在两年RSA迁移方面与经验证的对称设计相当,可作为临床选择依据之一。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: The objective of this study was to compare the two-year migration pattern and clinical outcomes of a total knee arthroplasty (TKA) with an asymmetrical tibial design (Persona PS) and a well-proven TKA with a symmetrical tibial design (NexGen LPS). METHODS: A randomized controlled radiostereometric analysis (RSA) trial was conducted including 75 cemented posterior-stabilized TKAs. Implant migration was measured with RSA. Maximum total point motion (MTPM), translations, rotations, clinical outcomes, and patient-reported outcome measures (PROMs) were assessed at one week postoperatively and at three, six, 12, and 24 months postoperatively. RESULTS: A linear mixed-effect model using RSA data of 31 asymmetrical and 38 symmetrical TKAs did not show a difference in mean MTPM migration pattern of the tibial or femoral components. Mean tibial component MTPM at two years postoperative of the asymmetrical TKA design was 0.93 mm and 1.00 mm for the symmetrical design. For the femoral component these values were 1.04 mm and 1.14 mm, respectively. No significant differences were observed in other migration parameters or in clinical and PROM measurements. CONCLUSION: The TKA design with an asymmetrical tibial component has comparable component migration with the proven TKA with a symmetrical tibial component. This suggests the risk of long-term aseptic loosening of the two designs is comparable. Cite this article: Bone Joint J 2021;103-B(5):855-863.

来源:Bone Joint J,查看原文。版权归原作者所有。