小头金属对金属全髋关节置换术的长期结局:15至22年随访
研究纳入1997至2004年168例Metasul金属对金属全髋关节置换术,平均随访15年。19髋翻修,其中12髋因召回髋臼组件,2髋为真正界面失效。除外召回组件,Metasul界面在相对年轻人群中表现极佳,界面相关翻修仅2例。
要点
- 召回髋臼组件是翻修主因,占19髋翻修中的12髋。
- 真正金属对金属界面失效仅2髋,界面生存率98.8%。
- 平均Harris髋关节评分85分,UCLA评分6分,钴铬离子水平中位数分别为1.0和0.91 μg/L。
核心结论
- 除外召回组件,Metasul关节界面在至少15年随访中表现极佳,整个队列仅2例界面相关翻修。
研究设计与样本
- 1997至2004年连续行168例初次Metasul全髋关节置换术,共144例患者,多数使用非骨水泥股骨柄和多孔涂层髋臼组件及28 mm头。
主要发现
- 168髋中19髋翻修,平均翻修时间15年,其中12髋因召回髋臼组件,5髋为非界面相关并发症,2髋为真正界面表面失效。
- 全因翻修生存分布函数为88.4%,承重界面相关为98.8%,召回髋臼杯为92.6%。
- 平均Harris髋关节评分85分,UCLA评分6分;血钴和铬水平中位数分别为1.0和0.91 μg/L。
关键数值与推荐
- 全因翻修生存率88.4%,承重界面相关翻修生存率98.8%,召回髋臼杯翻修生存率92.6%。
- 平均Harris髋关节评分85分,UCLA评分6分,血钴中位数1.0 μg/L,血铬中位数0.91 μg/L。
临床怎么用
- 对于使用Metasul界面的全髋关节置换术患者,需关注召回髋臼组件的长期随访和翻修风险。
- 在排除召回组件影响后,该金属对金属界面在相对年轻人群中长期耐久性良好,界面相关翻修率低。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The Metasul articular interface was a second-generation metal-on-metal (MoM) total hip arthroplasty (THA) that was introduced as a promising interface with improved manufacturing technology, better clearances, and enhanced metal hardness. In December 2001, the manufacturer recalled these implants due to the failure of cup osseointegration. METHODS: Between 1997 and 2004, 168 consecutive primary Metasul THAs were performed in 144 patients. Most patients received a cementless femoral and porous-coated acetabular component with 28 mm head. A competing risk analysis was performed for determination separately for bearing surface-related, recalled bearing failure, and end point revision for any reason. For clinical patient evaluation, we used Harris hip score and University of California at Los Angeles scores. Cobalt and chromium ion level measurement and standard radiographic assessment was performed. RESULTS: Of the 168 THAs, 19 hips were revised at a mean period of 15 years as follows: 12 of them were due to recalled acetabular component, five hips had noninterface-related complication, and two true interface surface failure. The survival distribution function of all hips with revision for any reason was 88.4%, for bearing interface relation 98.8%, and 92.6% for recalled cups. The mean Harris hip score and University of California at Los Angeles scores were 85 and 6, respectively, and the median Co and Cr blood levels were 1.0 and 0.91 μg/L. CONCLUSION: Excluding the recalled components, Metasul articular interface has performed extremely well at a minimum follow-up of 15 years in this relatively young population. There were two interface-related revisions in the entire cohort.
引用
- Maliarov A, Huk OL, Epure LM, et al. Long-Term Outcome of Small Head Metal-On-Metal Total Hip Arthroplasty: A 15-to-22 Year Follow-Up. J Arthroplasty. 2021. PMID: 33941413. doi:10.1016/j.arth.2021.04.011
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。