旋转铰链膝关节假体聚乙烯部件取回分析

Retrieval Analysis of Polyethylene Components in Rotating Hinge Knee Arthroplasty Implants.
PubMed 文献 · J Arthroplasty · 2021-04-13

BACKGROUND: This study examined a cohort of retrieved rotating hinge (RH) total knee arthroplasty implants of four different designs with emphasis on the surface damage observed on the polyethylene components.…

要点

  • Stryker Howmedica MRH组背面磨损最少但关节面磨损显著多于其他两组。
  • 除NexGen外,机械原因翻修和植入时间长的假体总损伤评分增加。

核心结论

  • 没有单一旋转铰链假体设计在最小化聚乙烯磨损损伤方面表现出优势。

主要发现

  • Stryker Howmedica MRH组背面磨损最少,但关节面磨损显著多于其他三组中的两组。
  • 除NexGen外,机械原因翻修和植入时间更长的假体总损伤评分增加。

临床怎么用

  • 临床选择旋转铰链假体时,无需仅基于聚乙烯磨损损伤的严重程度偏好某一种设计。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: This study examined a cohort of retrieved rotating hinge (RH) total knee arthroplasty implants of four different designs with emphasis on the surface damage observed on the polyethylene components. Our purpose was to determine if differences in polyethylene damage existed among the designs, and if those differences could be explained by differences in design characteristics. MATERIALS AND METHODS: Seventy-two RH implants from four manufacturers (DePuy LPS/SROM, Zimmer NexGen, Stryker Howmedica MRH, and Biomet Finn-OSS) removed at the time of revision performed between 2002 and 2017 were identified in our institutional retrieval registry. Damage to the surfaces of the polyethylene was assessed using a subjective grading system and evaluated in multiple zones. Design characteristics that were evaluated included the following: location of the dwell point on the polyethylene component, posterior position of the axle, and amount of hyperextension and rotation allowed by the implant. RESULTS: There were no differences in total damage scores between the four implant groups (P = .45). The Stryker Howmedica MRH group showed the least backside wear of all implants but significantly more articular-sided wear compared with two of the other three groups. All implants except NexGen showed increased total damage scores in implants revised for mechanical (vs nonmechanical) reasons and in implants with a longer duration of implantation. CONCLUSION: No single implant design emerged as superior in terms of minimizing polyethylene wear damage. Polyethylene damage existed in various locations but was not different in severity across designs, suggesting that there is no clear superior RH design that minimizes overall articular surface wear compared with other designs.

来源:J Arthroplasty,查看原文。版权归原作者所有。