文献速递关节外科

双动全髋假体股骨颈切迹:值得担忧吗?

Femoral Neck Notching in Dual Mobility Implants: Is This a Reason for Concern?
PubMed 文献 · J Arthroplasty · 2021-03-24

切迹最早于术后平均1.3年出现,深度为假体颈直径的1.7%-20%,平均随访2.7年未见假体柄断裂、脱位或翻修。

要点

  • 双动全髋假体股骨颈切迹发生率为3.5%,均见于模块化圆柱球形钴铬假体。
  • 切迹最早于术后平均1.3年出现,深度为假体颈直径的1.7%-20%。
  • 平均随访2.7年,切迹未导致假体柄断裂、脱位或翻修。

核心结论

  • 双动全髋假体股骨颈切迹发生率为3.5%,略高于该高风险队列的脱位率,但尚未导致灾难性失败。

研究设计与样本

  • 多中心回顾性研究,纳入256例模块化和32例一体式双动假体,至少1年临床和影像学随访,包括112例翻修、4例转换全髋和172例初次全髋。

主要发现

  • 切迹最早于术后平均1.3年(范围0.5-2.7年)出现,位于假体股骨颈前侧20%、上侧60%、后侧20%。
  • 切迹深度为假体颈直径的1.7%-20%,8例切迹患者存在可能影响脊柱骨盆活动的腰椎病变。
  • 平均随访2.7年(范围1-7.6年),无假体柄断裂,切迹患者无脱位或翻修。

临床怎么用

  • 使用模块化圆柱球形钴铬双动假体时,应关注股骨颈切迹风险,尤其对合并腰椎病变影响脊柱骨盆活动的患者。
  • 假体位置、脊柱骨盆活动和假体设计可能影响切迹风险,需进一步研究临床后遗症。

局限性

  • 回顾性设计,随访时间有限,切迹的长期临床后遗症尚不明确。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Dual mobility (DM) total hip arthroplasty (THA) implants have been advocated for patients at risk for impingement due to abnormal spinopelvic mobility. Impingement against cobalt-chromium acetabular bearings, however, can result in notching of titanium femoral stems. This study investigated the incidence of femoral stem notching associated with DM implants and sought to identify risk factors. METHODS: A multicenter retrospective study reviewed 256 modular and 32 monoblock DM components with minimum 1-year clinical and radiographic follow-up, including 112 revisions, 4 conversion THAs, and 172 primary THAs. Radiographs were inspected for evidence of femoral notching and to calculate acetabular inclination and anteversion. Revisions and dislocations were recorded. RESULTS: Ten cases of femoral notching were discovered (3.5%), all associated with modular cylindrospheric cobalt-chromium DM implants (P = .049). Notches were first observed radiographically at mean 1.3 years after surgery (range 0.5-2.7 years). Notch location was anterior (20%), superior (60%), or posterior (20%) on the prosthetic femoral neck. Notch depth ranged from 1.7% to 20% of the prosthetic neck diameter. Eight cases with notching had lumbar pathology that can affect spinopelvic mobility. None of these notches resulted in stem fracture, at mean 2.7-year follow-up (range 1-7.6 years). There were no dislocations or revisions in patients with notching. CONCLUSION: Femoral notching was identified in 3.5% of DM cases, slightly surpassing the dislocation rate in a cohort selected for risk of impingement and instability. Although these cases of notching have not resulted in catastrophic failures thus far, further study of clinical sequelae is warranted. Component position, spinopelvic mobility, and implant design may influence risk.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Lygrisse KA, Matzko C, Shah RP, et al. Femoral Neck Notching in Dual Mobility Implants: Is This a Reason for Concern? J Arthroplasty. 2021. PMID: 33875287. doi:10.1016/j.arth.2021.03.043

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。