全髋翻修假体锥度微动腐蚀:锥度几何、头尺寸与头材料的交互作用

Assessing Taper Geometry, Head Size, Head Material, and Their Interactions in Taper Fretting Corrosion of Retrieved Total Hip Arthroplasty Implants.
PubMed 文献 · J Arthroplasty · 2021-02-23

研究对157例取出的全髋关节置换假体按股骨头材料(ZTA vs CoCr)、头尺寸(32mm vs 36mm)和锥度几何(12/14 vs V40)分组评估锥度微动与腐蚀分级。析因分析显示头材料和锥度几何是显著影响因素,ZTA、32mm头配12/14锥度的中重度损伤率最低。

要点

  • ZTA头中重度微动腐蚀率为13%,低于CoCr头的38%。
  • 12/14锥度中重度损伤率为13%,低于V40锥度的35%。
  • CoCr头配V40锥度中重度损伤率最高(32mm为47%,36mm为59%)。

核心结论

  • ZTA材料、32mm头与12/14锥度组合的中重度微动腐蚀损伤率较低。
  • 析因分析显示头材料、锥度几何及其交互作用是损伤分级的最显著相关因素。

研究设计与样本

  • 对157例取出全髋关节置换假体进行锥度微动和腐蚀分级评估的病例系列研究(证据等级IV)。
  • 按头材料(ZTA 95例 vs CoCr 62例)、头尺寸(32mm 56例 vs 36mm 101例)、锥度几何(12/14 84例 vs V40 73例)定义8个队列,采用2^3析因模型统计分析。

主要发现

  • 中重度微动腐蚀率:ZTA头13% vs CoCr头38%。
  • 中重度微动腐蚀率:32mm头18% vs 36mm头26%。
  • 中重度微动腐蚀率:12/14锥度13% vs V40锥度35%。
  • ZTA+32mm头中重度损伤率最低(12/14锥度2%,V40锥度7%)。
  • CoCr头配V40锥度中重度损伤率最高(32mm 47%,36mm 59%)。

关键数值与推荐

  • 析因分析中与总微动和腐蚀损伤显著相关的因素为头材料和锥度几何。
  • 头材料-锥度几何交互作用对总腐蚀损伤也是显著因素。

临床怎么用

  • 选择ZTA头、32mm头尺寸与12/14锥度组合可能降低锥度微动腐蚀损伤风险。
  • 避免CoCr头与V40锥度组合,因其中重度损伤率最高。

局限性

  • 本研究为病例系列(证据等级IV),非随机对照设计。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Decreased fretting and corrosion damage at the taper interface of retrieved ceramic-on-polyethylene total hip arthroplasty (THA) implants has been consistently reported; however, resultant fretting corrosion as a function of femoral head size and taper geometry has not been definitively explained. METHODS: Eight cohorts were defined from 157 retrieved THA implants based on femoral head composition (n = 95, zirconia-toughened alumina, ZTA vs n = 62, cobalt-chromium alloy, CoCr), head size (n = 56, 32mm vs n = 101, 36mm), and taper geometry (n = 84, 12/14 vs n = 73, V40). THA implants were evaluated and graded for taper fretting and corrosion. Data were statistically analyzed, including via a 23 factorial modeling. RESULTS: Factorial-based analysis indicated the significant factors related to both resultant (summed) fretting and corrosion damage were head material and taper geometry; head material-taper geometry interaction was also a significant factor in resultant corrosion damage. Lower rates of moderate-to-severe fretting and corrosion damage were exhibited on ZTA heads (ZTA = 13%, CoCr = 38%), smaller heads (32mm = 18%, 36mm = 26%), and 12/14 tapers (12/14 = 13%, V40 = 35%). ZTA+32mm heads demonstrated the lowest rates of moderate-to-severe fretting and corrosion damage (12/14 = 2%, V40 = 7%), whereas CoCr heads with V40 tapers demonstrated the greatest rates of moderate-to-severe damage (32mm = 47%, 36mm = 59%). CONCLUSION: In this series, retrieved implants with ZTA, 32-mm heads paired with 12/14 tapers exhibited lower rates of moderate-to-severe damage. Factorial analysis showed head material, taper geometry, and their interactions were the most significant factors associated with resultant damage grades. Isolating implant features may provide additional information regarding factors leading to fretting and corrosion damage in THA. LEVEL OF EVIDENCE: IV (case series).