基于影像的机器人辅助系统在内侧单髁膝关节置换术中的尺寸准确性优于无影像和传统技术
本研究比较了传统技术、无影像机器人辅助系统和基于术前CT的影像机器人辅助系统在内侧单髁膝关节置换术中的尺寸准确性。每组93例,通过术后X线评估六个参数,错误尺寸定义为超过3mm的过大或过小。结果显示影像机器人系统显著降低胫骨和股骨后侧尺寸过小的风险。
要点
- 影像机器人组胫骨后侧尺寸过小率仅6.5%,显著低于传统组的47.3%。
- 股骨后侧尺寸过小率在影像机器人组为12.9%,传统组为30.1%。
- 传统组股骨后侧偏移增加更常见(43%),影像机器人组仅8.6%。
核心结论
- 机器人辅助技术可降低内侧单髁膝关节置换术中的尺寸过小风险,尤其是基于术前CT的影像系统。
研究设计与样本
- 三组各93例内侧单髁膝关节置换术,分别采用传统技术、无影像机器人辅助系统和基于术前CT的影像机器人辅助系统。
主要发现
- 传统组胫骨后侧尺寸过小率为47.3%(n=44),无影像组为29%(n=27),影像组为6.5%(n=6),p<0.001。
- 传统组股骨后侧尺寸过小率为30.1%(n=28),无影像组为7.5%(n=7),影像组为12.9%(n=12),p<0.001。
- 传统组股骨后侧偏移增加率为43%(n=40),无影像组为30.1%(n=28),影像组为8.6%(n=8),p<0.001。
- 所有组均无股骨或胫骨假体显著悬突。
关键数值与推荐
- 错误尺寸定义为尺寸过大或过小超过3mm。
- 通过术后X线片比较六个参数评估单髁膝关节置换术尺寸。
临床怎么用
- 对于内侧单髁膝关节置换术,采用基于术前CT的影像机器人辅助系统可能提高尺寸准确性,减少后侧尺寸过小和股骨后侧偏移增加。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIMS: Ideal component sizing may be difficult to achieve in unicompartmental knee arthroplasty (UKA). Anatomical variants, incremental implant size, and a reduced surgical exposure may lead to over- or under-sizing of the components. The purpose of this study was to compare the accuracy of UKA sizing with robotic-assisted techniques versus a conventional surgical technique. METHODS: Three groups of 93 medial UKAs were assessed. The first group was performed by a conventional technique, the second group with an image-free robotic-assisted system (Image-Free group), and the last group with an image-based robotic arm-assisted system, using a preoperative CT scan (Image-Based group). There were no demographic differences between groups. We compared six parameters on postoperative radiographs to assess UKA sizing. Incorrect sizing was defined by an over- or under-sizing greater than 3 mm. RESULTS: There was a higher rate of tibial under-sizing posteriorly in the conventional group compared to robotic-assisted groups (47.3% (n = 44) in conventional group, 29% (n = 27) in Image-Free group, 6.5% (n = 6) in Image-Based group; p < 0.001), as well as a higher rate of femoral under-sizing posteriorly (30.1% (n = 28) in conventional group, 7.5% (n = 7) in Image-Free group, 12.9% (n = 12) in Image-Based group; p < 0.001). The posterior femoral offset was more often increased in the conventional group, especially in comparison to the Image-Based group (43% (n = 40) in conventional group, 30.1% (n = 28) in Image-Free group, 8.6% (n = 8) in Image-Based group; p < 0.001). There was no significant overhang of the femoral or tibial implant in any groups. CONCLUSION: Robotic-assisted surgical techniques for medial UKA decrease the risk of tibial and femoral under-sizing, particularly with an image-based system using a preoperative CT scan. Cite this article: Bone Joint J 2021;103-B(4):610-618.
引用
- Batailler C, Bordes M, Lording T, et al. Improved sizing with image-based robotic-assisted system compared to image-free and conventional techniques in medial unicompartmental knee arthroplasty. Bone Joint J. 2021. PMID: 33789484. doi:10.1302/0301-620X.103B4.BJJ-2020-1453.R1
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。