椎体孔隙率与植入物载荷模式对腰椎全椎间盘置换后骨组织应力的影响

The Role of Vertebral Porosity and Implant Loading Mode on Bone-Tissue Stress in the Human Vertebral Body Following Lumbar Total Disc Arthroplasty.
PubMed 文献 · Spine (Phila Pa 1976) · 2021-01-01

基于12例人腰椎尸检micro-CT的有限元分析显示,全椎间盘置换植入物导致的骨组织最大应力增加依赖载荷模式,且与椎体骨体积分数(BV/TV)显著相关;高孔隙率椎体应力增加更明显,屈伸位撞击时更突出。

要点

  • TDA植入后骨组织最大应力增加在屈伸位撞击时大于轴向压缩。
  • BV/TV越低(孔隙率越高),植入导致的应力增加越显著。
  • 载荷模式与BV/TV存在交互作用,共同影响应力变化。

核心结论

  • 常规尺寸TDA植入物使高孔隙率椎体的骨组织应力不成比例增加,屈伸位撞击时更明显。

研究设计与样本

  • 基于12例人腰椎尸检micro-CT数据构建高分辨率有限元模型(单元边长37 μm),模拟椎间盘-椎体-植入物运动节段。

主要发现

  • 植入导致的骨组织最大应力增加取决于载荷模式(P < 0.001),屈伸位撞击时增加大于轴向压缩。
  • 最大应力变化与BV/TV显著相关(P = 0.002),高孔隙率椎体应力增加更显著。
  • 载荷模式与BV/TV存在显著交互作用(P = 0.002),载荷模式改变BV/TV与最大应力变化的关系。

关键数值与推荐

  • 12例腰椎(8男4女,51-89岁),BV/TV=0.060-0.145,植入物加载至800 N轴向压缩及屈伸位撞击。

临床怎么用

  • 对高孔隙率椎体行TDA时需关注植入物下沉风险,尤其屈伸位撞击载荷下。

局限性

  • 样本量小(12例),为离体有限元分析,结论需临床验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

STUDY DESIGN: Micro-computed tomography- (micro-CT-) based finite element analysis of cadaveric human lumbar vertebrae virtually implanted with total disc arthroplasty (TDA) implants. OBJECTIVE: (1) Assess the relationship between vertebral porosity and maximum levels of bone-tissue stress following TDA; (2) determine whether the implant's loading mode (axial compression vs. sagittal bending) alters the relationship between vertebral porosity and bone-tissue stress. SUMMARY OF BACKGROUND DATA: Implant subsidence may be related to the bone biomechanics in the underlying vertebral body, which are poorly understood. For example, it remains unclear how the stresses that develop in the supporting bone tissue depend on the implant's loading mode or on typical inter-individual variations in vertebral morphology. METHODS: Data from micro-CT scans from 12 human lumbar vertebrae (8 males, 4 females; 51-89 years of age; bone volume fraction [BV/TV] = 0.060-0.145) were used to construct high-resolution finite element models (37 μm element edge length) comprising disc-vertebra-implant motion segments. Implants were loaded to 800 N of force in axial compression, flexion-, and extension-induced impingement. For comparison, the same net loads were applied via an intact disc without an implant. Linear regression was used to assess the relationship between BV/TV, loading mode, and the specimen-specific change in stress caused by implantation. RESULTS: The increase in maximum bone-tissue stress caused by implantation depended on loading mode (P < 0.001), increasing more in bending-induced impingement than axial compression (for the same applied force). The change in maximum stress was significantly associated with BV/TV (P = 0.002): higher porosity vertebrae experienced a disproportionate increase in stress compared with lower porosity vertebrae. There was a significant interaction between loading mode and BV/TV (P = 0.002), indicating that loading mode altered the relationship between BV/TV and the change in maximum bone-tissue stress. CONCLUSION: Typically-sized TDA implants disproportionately increase the bone-tissue stress in more porous vertebrae; this affect is accentuated when the implant impinges in sagittal bending.Level of Evidence: N/A.