当代骨水泥型全膝关节置换术中固定的「薄弱环节」在哪里?
该回顾性分析纳入2005至2017年因无菌性胫骨松动接受翻修的149例患者,通过手术记录与影像学确定固定失败部位。结果显示失败更多见于假体-骨水泥界面(140/149,94.0%)而非骨水泥-骨界面(9/149,6.0%),提示假体-骨水泥界面是当代骨水泥型全膝关节置换胫骨固定的薄弱环节。
要点
- 无菌性胫骨松动翻修中94.0%(140/149)失败位于假体-骨水泥界面。
- 全部149例均有胫骨托前方异位骨形成,与失败界面无关。
核心结论
- 本研究中最常见的界面失败位于假体-骨水泥界面,为140/149(94.0%)。
研究设计与样本
- 回顾性分析2005至2017年单中心因无菌性胫骨松动接受翻修的149例患者,复习手术记录与X线片判定固定失败部位与模式。
主要发现
- 假体-骨水泥界面失败140/149(94.0%),骨水泥-骨界面失败9/149(6.0%)。
- 假体-骨水泥界面失败者中49/140(35.0%)为无成角的界面分离。
- 全部149例均存在胫骨托前方异位骨形成,且与失败界面无关。
临床怎么用
- 因失败主要位于假体-骨水泥界面,可考虑通过改进胫骨托下表面设计以增强假体与骨水泥的固定。
- 对过度活跃或体重指数升高的患者,改进胫骨托下表面特征可能有助于降低该界面失败风险。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Aseptic tibial loosening is a frequent cause of long-term failure following primary cemented total knee replacement. Failure of the tibial implant can occur at the implant-cement interface or at the cement-bone interface. Currently, it is unknown at which interface failure occurs in cases of aseptic tibial loosening. The following study was designed to determine which interface represents the "weak link" for tibial implant fixation. METHODS: We performed a retrospective analysis of 149 patients who were revised secondary to aseptic tibial loosening at our institution from 2005 to 2017. Operative reports and radiographs were reviewed on each patient to determine the location and pattern of fixation failure. RESULTS: Implant failure was more prevalent at the implant-cement than cement-bone interface, 140/149 (94.0%) vs 9/149 (6.0%). Additionally, we noted 2 distinct patterns of failure in patients that loosened at the implant-cement interface. Ninety of 140 (64.3%) patients developed varus collapse pattern of failure. Forty-nine of 140 (35.0%) patients developed failure between the implant-cement interface without angulation. All 149 patients had heterotopic bone formation anterior to the tibial baseplate, which was consistent regardless of which interface failed. CONCLUSION: The most frequent interface failure identified in our study was at the implant-cement interface, 140/149 (94.0%). This finding has substantial clinical ramifications. Because failure was predominantly at the implant-cement interface there may be design opportunities for increasing implant fixation to cement. Implants with improved undersurface tibial tray features may be necessary to mitigate the risk of failure at this interface, especially in overly active patients or those with elevated body mass indices.
引用
- Martin JR, Otero JE, Mason JB, et al. Where Is the "Weak Link" of Fixation in Contemporary Cemented Total Knee Replacements? J Arthroplasty. 2021. PMID: 33676813. doi:10.1016/j.arth.2021.02.029
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。