内侧平台衬垫形合度对PS全膝关节置换术中屈曲中期矛盾运动的影响
本研究利用经验证的计算模型,探究后稳定型(PS)全膝关节置换术中内侧平台衬垫矢状面形合度改变对屈曲中期矛盾运动的影响。结果显示,增加内侧形合度可减少屈曲中期内侧髁的前向滑动,但会增高股胫接触力,峰值达3.13倍体重。
要点
- 增加内侧形合度可减少PS TKA屈曲中期内侧髁前向滑动。
- 高形合度设计股胫接触力峰值3.13倍体重,高于其他设计3.0倍。
- 形合度增加可能引入运动学冲突,需权衡滑动与受力。
核心结论
- 增加衬垫内侧形合度可减少PS TKA屈曲中期前向滑动,但以增加股胫接触力为代价,可能源于高约束设计引入的运动学冲突。
研究设计与样本
- 使用经验证的计算模型,改变PS TKA内侧平台矢状面形合度,评估基线、增加和减少形合度三种场景的力学。
主要发现
- 从完全伸直到约70°屈曲,增加形合度设计内侧髁前向滑动极小,基线设计滑动2 mm,减少形合度设计滑动4 mm。
- 凸轮-立柱啮合后,三种设计内侧髁均持续后滚。
- 外侧髁在全部屈曲范围内三种设计均持续后滚。
- 增加形合度设计股胫接触力峰值3.13倍体重,其他两种设计为3.0倍体重。
临床怎么用
- 临床选择PS TKA衬垫时需权衡内侧形合度对屈曲中期稳定性和接触力的影响。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: One of the most common kinematic abnormalities reported for posterior-stabilized (PS) total knee arthroplasty (TKA) design is paradoxical anterior sliding during early and mid-flexion. PS TKAs have been designed such that the cam-post mechanism does not engage until later in flexion, making these implants vulnerable to anterior sliding during early and mid-flexion. The objective of this study is to investigate the biomechanical effect of increasing bearing conformity on a PS TKA. METHODS: Using a validated computational model of the knee joint, the sagittal conformity of the medial plateau of a PS TKA design was altered. Three scenarios were created and evaluated for mechanics: (1) baseline conformity, (2) increased conformity, and (3) decreased conformity. RESULTS: From full extension to approximately 70° of knee flexion, the medial condyle demonstrated minimal anterior sliding for the increased medial conformity design but revealed anterior sliding of 2 and 4 mm for the baseline and decreased conformity designs, respectively. After cam-post engagement, the medial condyle consistently rolled back for all 3 designs. The lateral condyle experienced consistent rollback throughout the entire flexion range for all 3 designs. However, femorotibial contact force was higher for the increased conformity design, peaking at 3.13 times body weight (×BW) compared to 3.0 × BW contact force for other 2 designs. CONCLUSION: Increasing medial conformity of the bearing insert appears to reduce mid-flexion sliding for PS TKA designs, although this comes at the expense of increased femorotibial forces. This could be due to kinematic conflicts that may be introduced with highly constraining designs.
引用
- Khasian M, Meccia BA, LaCour MT, et al. Effects of the Medial Plateau Bearing Insert Conformity on Mid-Flexion Paradoxical Motion in a Posterior-Stabilized Total Knee Arthroplasty Design. J Arthroplasty. 2021. PMID: 33602585. doi:10.1016/j.arth.2021.01.071