文献速递关节外科

高黏度骨水泥全膝关节置换后胫骨组件迁移研究

Tibial Component Migration After Total Knee Arthroplasty With High-Viscosity Bone Cement.
PubMed 文献 · J Arthroplasty · 2021-02-04

结果显示2年平均最大总点运动为0.41mm(SD 0.17),低于公认可接受迁移阈值,与中黏度骨水泥对照组无显著差异,Oxford 12膝关节评分从术前19(SD 7)改善至2年42(SD 8),提示无菌性松动风险低。

要点

  • 高黏度骨水泥固定胫骨组件2年迁移模式稳定,低于可接受阈值。
  • 迁移程度与中黏度骨水泥对照组无显著差异。
  • Oxford 12膝关节评分2年显著改善,临床效果良好。

核心结论

  • 高黏度骨水泥在2年随访中表现出可接受的迁移模式,提示无菌性松动风险低。

主要发现

  • 平均最大总点运动迁移1年时为0.40mm(SD 0.16),2年时为0.41mm(SD 0.17),呈稳定固定模式。
  • 迁移低于已发表的可接受迁移阈值,且与同一中心先前发表的中黏度骨水泥对照组无显著差异。
  • 1例假体在1至2年间出现持续迁移,但临床无症状。
  • 术后至少1年时平均最大总点运动诱导位移为0.3mm(SD 0.12)。
  • 平均Oxford 12膝关节评分从术前19(SD 7)改善至术后2年42(SD 8)。

关键数值与推荐

  • 迁移评估采用放射立体测量分析(RSA),比较已发表的可接受迁移阈值。
  • 临床评估采用Oxford 12膝关节评分和满意度记录。

临床怎么用

  • 高黏度骨水泥可用于全膝关节置换胫骨组件固定,短期随访显示固定稳定、无菌性松动风险低。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: High-viscosity (HV) bone cements have been formulated to offer potentially advantageous handling characteristics. However, alteration in the handling characteristics could influence implant fixation and survival. The primary objective of this study was to use radiostereometric analysis after total knee arthroplasty to assess the migration of the Triathlon tibial component fixed with HV cement (Simplex HV). METHODS: Twenty-three patients were followed for two years with radiostereometric analysis examinations at 6 visits. Migration was compared with published thresholds and with a control group from a previously published study from the same center using the same implants fixed with a medium viscosity cement. Inducible displacement was assessed, and Oxford 12 Knee Scores and satisfaction were recorded. RESULTS: Mean maximum total point motion migration reaching 0.40 mm (SD 0.16) at one year, and 0.41mm (SD 0.17) at two years, demonstrating a pattern of stable fixation, below published thresholds of acceptable migration, and not significantly different from the control group. One implant had continuous migration between 1 and 2 years but was clinically asymptomatic. Mean maximum total point motion inducible displacement measured at least one year postoperatively was 0.3 mm (SD 0.12). Mean Oxford 12 Knee Scores improved from 19 (SD 7) preoperatively to 42 (SD 8) 2 years postoperatively. CONCLUSIONS: The use of HV cement demonstrated an acceptable pattern of migration at 2 years, indicating low risk for aseptic loosening.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Laende EK, Richardson CG, Meldrum AR, et al. Tibial Component Migration After Total Knee Arthroplasty With High-Viscosity Bone Cement. J Arthroplasty. 2021. PMID: 33632580. doi:10.1016/j.arth.2021.01.081

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。