文献速递关节外科

维生素E注入高交联与中度交联聚乙烯在反式全髋关节置换中的五年RSA随机对照试验结果

Results of a randomized controlled trial with five-year radiostereometric analysis results of vitamin E-infused highly crosslinked versus moderately crosslinked polyethylene in reverse total hip arthroplasty.
PubMed 文献 · Bone Joint J · 2020-01-01

该前瞻性随机对照试验纳入68例反式混合全髋关节置换患者,比较维生素E注入高交联聚乙烯(VEPE)与中度交联聚乙烯(ModXLPE)骨水泥髋臼假体的五年聚乙烯磨损和假体稳定性。结果显示VEPE组近端磨损和总三维磨损均低于ModXLPE组,两组假体稳定性相似且五年内保持恒定。

要点

  • VEPE组五年近端磨损0.17 mm,低于ModXLPE组的0.20 mm(p=0.005)。
  • 两组年近端磨损率低,分别为0.03 mm/年和0.04 mm/年。
  • 两组髋臼假体稳定性相似,五年内保持恒定。

核心结论

  • 五年时VEPE组聚乙烯磨损低于ModXLPE组,两组骨水泥假体年磨损率均低且假体稳定性相似并保持恒定。

研究设计与样本

  • 前瞻性随机对照试验,纳入68例反式混合全髋关节置换患者,随机分配至VEPE或ModXLPE骨水泥髋臼假体组,采用放射立体照相分析(RSA)评估五年聚乙烯磨损和髋臼假体稳定性(主要结局)。

主要发现

  • 五年时近端聚乙烯磨损VEPE组为0.17 mm(SD 0.15),ModXLPE组为0.20 mm(SD 0.09)(p=0.005)。
  • 年近端磨损率VEPE组为0.03 mm/年,ModXLPE组为0.04 mm/年。
  • 总三维磨损VEPE组为0.21 mm(SD 0.26),ModXLPE组为0.23 mm(SD 0.10)(p=0.009)。
  • 总三维髋臼杯平移VEPE组为0.72 mm(SD 0.70),ModXLPE组为0.50 mm(SD 0.44)(p=0.409)。

临床怎么用

  • 在反式混合全髋关节置换中,VEPE骨水泥髋臼假体五年磨损低于ModXLPE,假体稳定性相似。

局限性

  • 这些结果是否等同于更低的长期翻修率仍未知。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: Vitamin E-infused highly crosslinked polyethylene (VEPE) has been introduced into total hip arthroplasty (THA) with the aim of further improving the wear characteristics of moderately and highly crosslinked polyethylenes (ModXLPE and HXLPE). There are few studies analyzing the outcomes of vitamin E-infused components in cemented arthroplasty, though early acetabular component migration has been reported. The aim of this study was to measure five-year polyethylene wear and acetabular component stability of a cemented VEPE acetabular component compared with a ModXLPE cemented acetabular component. METHODS: In a prospective randomized controlled trial (RCT), we assessed polyethylene wear and acetabular component stability (primary outcome) with radiostereometric analysis (RSA) in 68 patients with reverse hybrid THA at five years follow-up. Patients were randomized to either a VEPE or a ModXLPE cemented acetabular component. RESULTS: Mean polyethylene wear in the proximal direction was 0.17 mm (SD 0.15) for the VEPE group and 0.20 mm (SD 0.09) for the ModXLPE group (p = 0.005) at five years. Annual proximal wear rates were 0.03 mm/year (VEPE) and 0.04 mm/year (ModXLPE). Total 3D wear was 0.21 mm (SD 0.26) and 0.23 mm (SD 0.10) for the VEPE and ModXLPE groups, respectively (p = 0.009). Total 3D cup translation was 0.72 mm (SD 0.70) (VEPE) and 0.50 mm (SD 0.44) (ModXLPE) (p = 0.409). CONCLUSION: At five years, there was less polyethylene wear in the VEPE group than in the ModXLPE group. Both VEPE and ModXLPE cemented components showed low annual wear rates. Component stability was similar in the two groups and remained constant up to five years. Whether these results will equate to a lower long-term revision rate is still unknown. Cite this article: Bone Joint J 2020;102-B(12):1646-1653.

来源:Bone Joint J,查看原文。版权归原作者所有。

引用

  1. Thoen PS, Nordsletten L, Pripp AH, et al. Results of a randomized controlled trial with five-year radiostereometric analysis results of vitamin E-infused highly crosslinked versus moderately crosslinked polyethylene in reverse total hip arthroplasty. Bone Joint J. 2020. PMID: 33249906. doi:10.1302/0301-620X.102B12.BJJ-2020-0721.R1