全髋关节置换中金属柄与陶瓷内衬的撞击:颈部并非唯一发生部位

Impingement Between the Metal Stem and the Ceramic Liner in Total Hip Arthroplasty: The Neck Is Not the Only Place Where Impingement Occurs.
PubMed 文献 · J Bone Joint Surg Am · 2021-01-01

本研究回顾性分析244例(197例患者)陶瓷对陶瓷 bearings 的全髋关节置换病例,随访≥15年,通过X线片评估柄上凹痕(U形磨损痕)分析撞击模式。结果显示23.4%病例出现凹痕,位于柄颈或柄肩,柄肩凹痕仅见于短颈头病例,且凹痕组杯外倾角更低、前倾角更高。结论提示柄肩与陶瓷内衬撞击可导致内衬骨折,建议避免蹲坐或盘腿坐地。

要点

  • 陶瓷对陶瓷全髋置换中,柄与陶瓷内衬撞击可发生于柄颈和柄肩。
  • 柄肩凹痕仅见于短颈头病例,占短颈病例的20.0%。
  • 凹痕组杯外倾角更低(p=0.01)、前倾角更高(p=0.01)。

核心结论

  • 陶瓷对陶瓷全髋关节置换中,相当比例患者发生柄与陶瓷内衬撞击,不仅位于颈部也位于柄肩。
  • 柄肩的猛烈突然撞击可导致陶瓷内衬骨折,设计柄时应考虑柄肩与内衬的撞击。
  • 建议患者尽量避免蹲坐或盘腿坐地。

研究设计与样本

  • 通过X线片重点评估柄上凹痕(U形磨损痕)和杯位置。

主要发现

  • 57例(23.4%)在77个部位检出凹痕,首次出现于术后8个月至14.8年。
  • 凹痕位于柄颈或柄肩,柄肩凹痕仅见于短颈头病例,29例(占短颈病例20.0%)。
  • 凹痕组杯外倾角低于无凹痕组(p=0.01),前倾角高于无凹痕组(p=0.01)。
  • 共5例陶瓷头骨折,其中1例翻修术后因假体柄肩撞击发生另一陶瓷内衬骨折。

关键数值与推荐

  • 凹痕检出率23.4%(57/244例)。
  • 柄肩凹痕在短颈病例中占20.0%(29例)。
  • 杯外倾角组间比较p=0.01,前倾角组间比较p=0.01。
  • 证据等级:治疗性IV级。

临床怎么用

  • 全髋关节置换术后应关注柄肩与陶瓷内衬的撞击风险,尤其使用短颈头时。
  • 影像随访时注意识别柄上凹痕,凹痕出现提示撞击可能。
  • 建议患者避免蹲坐或盘腿坐地等极限体位以减少撞击。

局限性

  • 本研究为回顾性设计,证据等级为治疗性IV级。
  • 仅纳入随访≥15年的病例,可能存在选择偏倚。
  • 凹痕的影像学识别可能受投照角度影响。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: It is assumed that impingement between the ceramic liner and the stem increases the possibility of a liner fracture in total hip arthroplasty with a ceramic-on-ceramic bearing. The purpose of this study was to analyze the pattern of the impingement by evaluating the notches (U-shaped indented wear scars engraved on the stem) on radiographs to determine when and where impingement develops and to analyze the factors affecting its occurrence. METHODS: Among the primary total hip arthroplasty cases using a ceramic-on-ceramic bearing performed from November 1997 to December 2003, 244 cases of 197 patients (123 male patients and 74 female patients) that had follow-up of ≥15 years were included. All of the radiographs were examined with special regard to the notches and the cup positions. RESULTS: Notches were detected at 77 sites of 57 cases (23.4%) for the first time between 8 months and 14.8 years after the surgical procedure. They were located on the neck or the shoulder of the stem. Shoulder notches were detected only in the cases with a short-neck head. Shoulder notches were found in 29 cases (20.0% of short-neck cases). Cup inclination was lower (p = 0.01) and anteversion was higher (p = 0.01) in the group with notches than the group without notches. There were 5 cases of ceramic head fracture. One of them experienced another ceramic liner fracture, assumed to be caused by prosthetic shoulder impingement, after the revision surgical procedure. CONCLUSIONS: The results of this study suggest that impingement between the stem and the ceramic liner occurs in a considerable proportion of patients who underwent total hip arthroplasty not only on the neck but also on the shoulder of the stem. Forceful and abrupt impingement on the stem shoulder can cause ceramic liner fracture. Impingement between the stem shoulder and the ceramic liner should be considered in designing a stem. It seems to be prudent to recommend that patients avoid squatting or sitting cross-legged on the floor as much as possible. LEVEL OF EVIDENCE: Therapeutic Level IV. See Instructions for Authors for a complete description of levels of evidence.