透视图像畸变是否影响直接前入路全髋关节置换术的肢体长度测量
该研究回顾74例直接前入路全髋关节置换术的术中透视骨盆图像,由两名独立读者测量图像周边垂直畸变并求和,模拟肢体长度侧侧对比。平均合并畸变为10.0mm(范围2.0-20.0mm),不同手术室间畸变有显著差异(P < .001),畸变与患者人口学或透视时间无关联(均P > .05)。结论认为透视畸变不可预测,在直接前入路全髋关节置换术中比较肢体长度时可能造成显著误差。
要点
- 直接前入路全髋关节置换术中透视图像平均合并垂直畸变达10.0mm(范围2.0-20.0mm)。
- 不同手术室间平均畸变存在显著差异(P < .001),畸变不可预测。
- 畸变与患者人口学特征或透视时间无显著关联(均P > .05)。
核心结论
- 透视畸变不可预测,在直接前入路全髋关节置换术中比较肢体长度时可能造成显著误差。
- 若未考虑该误差,可能导致非预期的假体位置和肢体长度差异。
研究设计与样本
- 回顾性分析74例直接前入路全髋关节置换术的术中透视骨盆图像,由两名独立读者测量。
- 所有图像使用同一台透视C臂机,并在影像增强器上附加不透射线网格。
主要发现
- 平均合并垂直畸变为10.0mm(范围2.0-20.0mm)。
- 不同手术室间平均畸变存在显著差异(P < .001)。
- 畸变与患者人口学特征或透视时间无关联(均P > .05)。
临床怎么用
- 在直接前入路全髋关节置换术中利用透视比较肢体长度时,应意识到图像畸变可能带来约10.0mm量级的误差。
- 不同手术室的透视畸变程度可能不同,需谨慎解读术中肢体长度测量。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Distortion is an intrinsic phenomenon associated with image-intensified fluoroscopy that is both poorly understood and infrequently appreciated by orthopedic surgeons. Little information exists regarding its potential influence on intraoperative parameters during orthopedic surgery, let alone during direct anterior (DA) total hip arthroplasty (THA). The purpose of this study was to quantify the amount of potential error caused by fluoroscopic distortion during DA THA. METHODS: Intra-operative fluoroscopic pelvic images from 74 DA THAs were reviewed by two independent readers. All images were obtained using the same fluoroscopic C-arm unit with a radiopaque grid attached to the image intensifier. The vertical distortion from a straight central horizontal line at the peripheries of images were measured and summed to yield the combined vertical distortion similar to how a surgeon calculates a side to side comparison of limb lengths. Simple linear regression was used to evaluate associations between total distortion and patient demographics, operating theaters, and various operative parameters. RESULTS: The average combined distortion was 10.0mm (range 2.0-20.0mm). There was a significant difference in the average distortion observed in different theaters (P < .001). There was no association between distortion and patient demographics or fluoroscopic time (all, P > .05). CONCLUSION: Fluoroscopic distortion is unpredictable and can cause a substantial amount of error when comparing limb lengths during DA THA. This is a critical finding as this amount of inaccuracy could lead to unintended implant positioning and limb-length discrepancies if unaccounted for.
引用
- Carlson VR, Elliott IS, DeKeyser GJ, et al. Are We Being Fooled by Fluoroscopy? Distortion May Affect Limb-Length Measurements in Direct Anterior Total Hip Arthroplasty. J Arthroplasty. 2020. PMID: 33303328. doi:10.1016/j.arth.2020.11.021
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。