髋关节表面置换至少13年随访的金属碎屑不良反应发生率评估
该回顾性研究纳入2000年1月至2005年8月间接受髋关节表面置换的连续患者,至少随访13年,评估金属碎屑不良反应(ARMD)的发生率。共102例患者123髋,7例8髋翻修,其中5髋因ARMD翻修,MR显示34%存在ARMD。尽管植入物生存率可接受,但ARMD发生率高,应告知患者相关风险并长期监测。
要点
- 髋关节表面置换至少13年随访中,MR显示34%的髋存在ARMD。
- 123髋中8髋翻修,5髋因ARMD翻修,2髋骨折,1髋股骨头坏死。
- 最佳情况下13年生存率为93.2%(109髋)。
核心结论
- 髋关节表面置换的植入物生存率可接受,但ARMD发生率高达34%。
- 医生和考虑髋关节表面置换的患者应知晓ARMD风险,以便知情选择并长期筛查。
研究设计与样本
- 回顾性研究,纳入2000年1月1日至2005年8月1日接受髋关节表面置换的连续患者,来自医院金属对金属髋关节置换监测数据库。
- 共102例患者123髋,每年随访症状和血离子水平,并行MR评估ARMD。
主要发现
- 7例患者8髋翻修:2髋骨折,1髋股骨头坏死,5髋ARMD。
- 最佳情况下13年生存率为93.2%(109/123髋)。
- MRI显示34%的髋存在ARMD。
临床怎么用
- 对髋关节表面置换患者应长期监测ARMD,包括症状、血离子水平和MR。
- 术前应告知患者ARMD风险及可能需翻修。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Hip Resurfacing (HR), although reducing in popularity, is still used in the younger male population. Excellent medium-term results have been published; however, the use of metal on metal has reduced with increased awareness of adverse reactions to metal debris (ARMD). ARMD has been shown to often be clinically "silent" following large Head MoM total hip replacement (THR). The purpose of our study was to report the incidence of ARMD following HR with a minimum follow-up of 13 years. METHODS: We performed a retrospective study of a consecutive series of patients who underwent HR between January 1, 2000 and August 1, 2005. All patients were entered into our hospital MoM hip replacement surveillance program database. Patients were reviewed yearly for symptoms and blood ion levels. Patients had Magnetic Resonance (MR) imaging to assess for ARMD. RESULTS: A total of 102 patients with 123 hip replacements were included in the study. Eight hips in 7 patients were revised: two for fracture, one for avascular necrosis, and five for ARMD. A best-case scenario of 109 (93.2%) resurfacings were surviving at 13 years. With regard to the radiological analysis, 34% were found to have ARMD on MR. CONCLUSION: While the implant survivorship in our series is acceptable, we found a high incidence of ARMD. Surgeons and patients with or considering a HR should be aware of the risk of ARMD developing. This allows an informed choice as to the best implant for their personal requirement and informs of the potential modes of failure and need for long-term screening.
引用
- Hastie GR, Collinson SC, Aqil A, et al. Study to Assess the Rate of Adverse Reaction to Metal Debris in Hip Resurfacing at a Minimum 13-year Follow-up. J Arthroplasty. 2020. PMID: 33183915. doi:10.1016/j.arth.2020.09.056
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。