文献速递关节外科

维生素E注入高交联聚乙烯未减少全膝关节置换术后体内磨损颗粒数量

Vitamin E-infused highly cross-linked polyethylene did not reduce the number of in vivo wear particles in total knee arthroplasty.
PubMed 文献 · Bone Joint J · 2020-01-01

本研究对比了维生素E注入高交联聚乙烯(E1)与常规聚乙烯(ArCom)在全膝关节置换术(TKA)后体内磨损颗粒的生成情况。共纳入34膝(每组17膝),术后平均3.4年获取滑液,用扫描电镜分析磨损颗粒。结果显示E1组颗粒总数更多、尺寸更小,表明E1在早期临床阶段未减少体内磨损颗粒数量。

要点

  • E1组体内聚乙烯磨损颗粒总数高于ArCom组(p=0.001)。
  • E1组颗粒尺寸小于ArCom组(p=0.001),纵横比也更小。
  • E1在早期临床阶段未减少体内磨损颗粒数量,需进一步随访。

核心结论

  • E1聚乙烯与ArCom相比,在早期临床阶段未减少体内聚乙烯磨损颗粒数量。

研究设计与样本

  • 共纳入34膝行TKA,ArCom组17膝,E1组17膝,除聚乙烯衬垫材料外假体相同。

主要发现

  • E1组磨损颗粒总数均值为6.9(SD 4.0)×10^7 counts/knee,高于ArCom组的2.2(SD 2.6)×10^7 counts/knee(p=0.001)。
  • E1组颗粒等效圆直径均值为0.5 μm(SD 0.1),小于ArCom组的1.5 μm(SD 0.3)(p=0.001)。
  • E1组颗粒纵横比均值为1.3(SD 0.1),小于ArCom组的1.4(SD 0.1)(p<0.001)。

临床怎么用

  • 对于新引入的E1材料在TKA中的应用,应进行仔细的长期随访以评估体内磨损表现。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: Vitamin E-infused highly cross-linked polyethylene (E1) has recently been introduced in total knee arthroplasty (TKA). An in vitro wear simulator study showed that E1 reduced polyethylene wear. However there is no published information regarding in vivo wear. Previous reports suggest that newly introduced materials which reduce in vitro polyethylene wear do not necessarily reduce in vivo polyethylene wear. To assist in the evaluation of the newly introduced material before widespread use, we established an in vivo polyethylene wear particle analysis for TKA. The aim of this study was to compare in vivo polyethylene wear particle generation between E1 and conventional polyethylene (ArCom) in TKA. METHODS: A total of 34 knees undergoing TKA (17 each with ArCom or E1) were investigated. Except for the polyethylene insert material, the prostheses used for both groups were identical. Synovial fluid was obtained at a mean of 3.4 years (SD 1.3) postoperatively. The in vivo polyethylene wear particles were isolated from the synovial fluid using a previously validated method and examined by scanning electron microscopy. RESULTS: The total number of polyethylene wear particles obtained from the knees with E1 (mean 6.9, SD 4.0 × 107 counts/knee) was greater than that obtained from those with ArCom (mean 2.2, SD 2.6 × 107 counts/knee) (p = 0.001). The particle size (equivalent circle of diameter) from the knees with E1 was smaller (mean 0.5 μm, SD 0.1) than that of knees with ArCom (mean 1.5, SD 0.3 μm) (p = 0.001). The aspect ratio of particles from the knees with E1 (mean 1.3, SD 0.1) was smaller than that with ArCom (mean 1.4, SD 0.1) (p < 0.001 ). CONCLUSION: This is the first report of in vivo wear particle analysis of E1. E1 polyethylene did not reduce the number of in vivo polyethylene wear particles compared with ArCom in early clinical stage. Further careful follow-up of newly introduced E1 for TKA should be carried out. Cite this article: Bone Joint J 2020;102-B(11):1527-1534.

来源:Bone Joint J,查看原文。版权归原作者所有。