氧化锆部件在髋关节脱位后除外的关节面保持光滑
研究分析了94个翻修取出的全髋和全膝关节假体股骨部件,采用光学显微镜、非接触式轮廓仪和扫描电镜观察关节面。结果显示氧化锆部件除髋关节脱位后外关节面保持光滑,4个因脱位翻修的氧化锆股骨头中3个出现严重损伤,陶瓷层剥脱且粗糙度达原始对照的23-32倍。
要点
- 氧化锆部件除髋关节脱位后外关节面保持光滑。
- 因脱位翻修的氧化锆股骨头中3/4出现陶瓷层剥脱和金属基底暴露。
- 因脱位翻修的氧化锆股骨头中位Sa=0.431 μm,钴铬股骨头为0.020 μm,P=.03。
核心结论
- 氧化锆股骨头脱位后陶瓷表面层剥脱和关节面显著粗糙并不罕见,氧化锆对髋关节脱位的耐受性远低于钴铬。
研究设计与样本
- 对94个翻修取出的全髋和全膝关节假体股骨部件(43个氧化锆全膝关节、21个氧化锆全髋关节、30个钴铬全髋关节)的关节面进行分析,采用光学显微镜、非接触式轮廓仪和扫描电子显微镜。
主要发现
- 4个因脱位翻修的氧化锆股骨头中3个出现关节面严重损伤,包括宏观陶瓷层剥脱和下方金属基底暴露,其粗糙度为原始氧化锆对照的23-32倍。
- 因脱位翻修时,氧化锆股骨头的中位Sa=0.431 μm,钴铬股骨头为0.020 μm,P=.03。
- 钴铬股骨头无论是否脱位整体粗糙度均较低,脱位者中位Sa=0.020 μm,非脱位者中位Sa=0.008 μm,P=.09(钴铬脱位者 vs 非脱位者)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Oxidized zirconium (OxZr) offers theoretical advantages in total hip and knee arthroplasty (THA and TKA, respectively) relative to other biomaterials by combining the tribological benefits of ceramics with the fracture toughness of metals. Yet, some studies have found that OxZr does not improve outcomes or wear rates relative to traditional bearing materials such as cobalt-chromium (CoCr). Separately, effacement of the thin ceramic surface layer has been reported for OxZr components, though the prevalence and sequelae are unclear. METHODS: To elucidate the in vivo behavior of OxZr implants, the articular surfaces of 94 retrieved THA and TKA femoral components (43 OxZr TKA, 21 OxZr THA, 30 CoCr THA) were analyzed using optical microscopy, non-contact profilometry, and scanning electron microscopy. RESULTS: We found that OxZr components maintain a smooth articular surface except following hip dislocation. Three of four OxZr femoral heads revised following dislocation exhibited severe damage to the articular surface, including macroscopic regions of ceramic-layer effacement and exposure of the underlying metal substrate; these components were 23-32 times rougher than pristine OxZr controls. When revised for dislocation, OxZr femoral heads were substantially rougher than CoCr femoral heads (median Sa = 0.431 v. 0.020 μm, P = .03). In contrast, CoCr femoral heads exhibited low overall roughness values regardless of whether they dislocated (median Sa = 0.020 v. 0.008 μm, P = .09, CoCr dislocators v. non-dislocators). CONCLUSIONS: Effacement of the ceramic surface layer and substantial articular surface roughening is not atypical following dislocation of OxZr femoral heads, making OxZr much less tolerant than CoCr to hip dislocation.
引用
- Bonnheim NB, Van Citters DW, Ries MD, et al. Oxidized Zirconium Components Maintain a Smooth Articular Surface Except Following Hip Dislocation. J Arthroplasty. 2020. PMID: 33246784. doi:10.1016/j.arth.2020.10.054
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。