文献速递关节外科

全膝关节置换术流程映射:器械设计的比较

Process Mapping Total Knee Arthroplasty: A Comparison of Instrument Designs.
PubMed 文献 · J Arthroplasty · 2020-09-23

该研究对来自10家公司、覆盖美国90%以上全膝关节置换术的17套膝关节植入系统进行流程映射,比较股骨、胫骨和髌骨准备所需器械步骤数。结果显示不同系统步骤数差异较大,流程映射可识别器械改进的潜在环节。

要点

  • 后稳定型膝关节系统所需步骤数多于保留交叉韧带型系统。
  • 不同制造商系统的器械步骤数存在差异,最少步骤的系统可为改进提供参考。
  • 流程映射可能有助于减少全膝关节置换术中的不必要步骤。

核心结论

  • 流程映射在所有研究的系统中识别出器械改进的潜在环节,提示可能在全膝关节置换术行业广泛减少手术步骤。

研究设计与样本

  • 评估来自10家公司、代表美国90%以上全膝关节置换术的17套膝关节植入系统,比较股骨、胫骨和髌骨准备所需器械,基于标准化手术流程并依据制造商推荐步骤调整,统计手术技师组装步骤、器械传递和外科医生步骤数。

主要发现

  • 保留交叉韧带型膝关节器械在研究系统中需要158至225个离散步骤。
  • 后稳定型膝关节需要181至230个步骤。
  • 保留交叉韧带型和后稳定型膝关节系统分别可改进至145步和163步。
  • 保留交叉韧带型系统中,Arthrex iBalance和Biomet Vanguard Microplasty所需步骤最少。
  • 后稳定型系统中,OrthoDevelopment Balanced和Corin Unity所需步骤最少。

局限性

  • 未按系统评估患者结局。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Total knee arthroplasty (TKA) is commonly performed with proprietary, manual instrumentation provided by the surgical implant manufacturer. Registry studies and meta-analysis, with few outliers, have consistently shown similar functional outcomes and implant survival after TKA regardless of implant manufacturer, implant design, or surgical technique. We hypothesized that process mapping could identify areas for improvement in TKA instrumentation. METHODS: Seventeen TKA implant systems from 10 companies representing over 90% of all TKAs performed in the United States were evaluated. Instrumentation required for femoral, tibial, and patellar preparation was compared. The number of steps including surgical technician assembly steps, instrument handoffs, and surgeon steps were tabulated based off application of a standardized surgical flow, adjusted for manufacturer-recommended steps during completion of a TKA operation. RESULTS: Cruciate-retaining (CR) knee instrumentation in studied systems required 158-225 discrete steps and posterior-stabilized (PS) knees required 181-230 steps. With the fewest steps for femoral, tibial, and patellar instrumentation, CR and PS knee systems could be improved to 145 and 163 steps, respectively. The Arthrex iBalance and the Biomet Vanguard Microplasty required fewest steps among CR systems; the OrthoDevelopment Balanced and the Corin Unity required fewest steps among PS systems. CONCLUSIONS: Process mapping identified potential areas for improved instrumentation in all studied systems, suggesting the possibility to reduce operative steps broadly across the TKA industry. Patient outcomes were not evaluated by system. Future implant system design changes may do well to reduce unnecessary steps and instrumentation.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Slick GS, Davis Iii CM, Elfar JC, et al. Process Mapping Total Knee Arthroplasty: A Comparison of Instrument Designs. J Arthroplasty. 2020. PMID: 33139131. doi:10.1016/j.arth.2020.09.035