全髋关节置换术后股骨假体设计对股骨近端骨量的影响:一项随机对照试验

Influence of Femoral Component Design on Proximal Femoral Bone Mass After Total Hip Replacement: A Randomized Controlled Trial.
PubMed 文献 · J Bone Joint Surg Am · 2021-01-01

这项随机对照试验比较了短柄近端多孔涂层Tri-Lock保骨柄与常规全涂层Corail假体在术后2年内的骨重塑和骨转换。46例接受Tri-Lock假体,40例接受Corail假体。结果显示两组均在股骨距和股骨近端外侧出现假体周围骨丢失(p<0.05),但常规假体组的骨密度降低幅度小于保骨假体组(p<0.001),该效应在股骨距和大转子区域最明显。两组骨转换标志物变化、PROM评分改善及不良事件率相似(p>0.05)。

要点

  • 两组均在股骨距和股骨近端外侧出现假体周围骨丢失(p<0.05)。
  • 常规Corail假体组骨密度降低幅度小于Tri-Lock保骨假体组(p<0.001)。
  • 两组骨转换标志物变化、PROM评分改善及不良事件率相似(p>0.05)。

核心结论

  • 本随机对照试验表明,旨在保留股骨近端骨的假体在这方面不一定优于常规非骨水泥设计。
  • DXA-RFA是检测假体周围骨重塑空间复杂模式的敏感工具。

研究设计与样本

  • 随机对照试验,46例接受Tri-Lock保骨柄,40例接受Corail假体,术后随访2年,在26周、52周和104周测量结局。

主要发现

  • 两组均在股骨距和股骨近端外侧出现假体周围骨丢失(p<0.05)。
  • 常规假体组骨密度降低幅度小于保骨假体组(p<0.001),该效应在股骨距和大转子区域最明显。
  • 部分区域可见骨密度小幅增加,常规假体组增加幅度大于保骨假体组(p<0.001)。
  • 两组骨转换标志物变化和PROM评分改善相似(p>0.05)。
  • 两组不良事件率相似(p>0.05)。

关键数值与推荐

  • 骨丢失采用高灵敏度双能X线吸收法无区域分析(DXA-RFA)测量。
  • 证据等级:治疗性I级。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: In this randomized controlled trial (RCT), we compared bone remodeling and bone turnover between 2 total hip arthroplasty implants-the short, proximally porous-coated Tri-Lock Bone-Preservation Stem and a conventional, fully-coated Corail prosthesis-over a 2-year postoperative period. METHODS: Forty-six participants received the Tri-Lock prosthesis and 40 received the Corail prosthesis. At baseline, the 2 groups had similar demographics, proximal femoral bone mineral density (BMD), bone turnover markers, radiographic canal flare index, and patient-reported outcome measure (PROM) scores. Outcomes were measured at weeks 26, 52, and 104. RESULTS: Loss of periprosthetic bone, measured by high-sensitivity dual x-ray absorptiometry region-free analysis (DXA-RFA), was identified at the calcar and proximal-lateral aspect of the femur in both prosthesis groups (p < 0.05). However, the conventional prosthesis was associated with a smaller reduction in BMD compared with the bone-preservation prosthesis (p < 0.001). This effect was most prominent in the region of the femoral calcar and greater trochanter. A small gain in BMD was also identified in some areas, and this gain was greater with the conventional than the bone-preservation prosthesis (p < 0.001). The 2 groups had similar changes in bone turnover markers and improvement in PROM scores over the study period (p > 0.05). The adverse-event rate was also similar between the groups (p > 0.05). CONCLUSIONS: This RCT shows that prostheses intended to preserve proximal femoral bone do not necessarily perform better in this regard than conventional cementless designs. DXA-RFA is a sensitive tool for detecting spatially complex patterns of periprosthetic bone remodeling. LEVEL OF EVIDENCE: Therapeutic Level I. See Instructions for Authors for a complete description of levels of evidence.

来源:J Bone Joint Surg Am,查看原文。版权归原作者所有。

引用

  1. Slullitel PA, Mahatma MM, Farzi M, et al. Influence of Femoral Component Design on Proximal Femoral Bone Mass After Total Hip Replacement: A Randomized Controlled Trial. J Bone Joint Surg Am. 2021. PMID: 33079901. doi:10.2106/JBJS.20.00351

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。