传感器引导膝关节平衡与传统平衡技术在全膝关节置换术中的随机对照试验
该前瞻性双盲随机对照试验纳入152例择期初次全膝关节置换术患者,比较传感器引导平衡与传统间隙平衡技术。传感器组术后冠状面间室压力差>15磅的不平衡比例显著低于对照组(5.3% vs 36%,P<.0001),但1年随访时膝关节屈曲度、Knee Society评分、Oxford评分及患者满意度均无显著差异。
要点
- 传感器引导平衡组术后不平衡比例显著低于传统间隙平衡组(5.3% vs 36%,P<.0001)。
- 1年随访时两组膝关节屈曲度、Knee Society评分及Oxford评分无显著差异。
- 1年满意度两组相当,对照组与实验组分别有81%报告非常满意(P=.992)。
核心结论
- 尽管传感器引导膝关节平衡装置在评估软组织包膜时提供了额外的定量反馈,但与传统间隙平衡技术相比,未能证明可改善临床结局或患者满意度。
研究设计与样本
- 前瞻性双盲随机对照试验,纳入择期初次全膝关节置换术患者152例,对照组76例,传感器引导实验组76例。
主要发现
- 对照组36%(27/76)膝关节基于平均冠状面间室压力差>15磅判定为不平衡,实验组仅5.3%(4/76)(P<.0001)。
- 术后1年两组屈曲度、Knee Society评分及Oxford评分无显著差异。
- 术后1年满意度两组相当,对照组与实验组分别有81%报告非常满意(P=.992)。
关键数值与推荐
- 传感器引导实验组调整目标为达到膝关节平衡,即间室压力差在15磅以内。
- 不平衡判定标准为平均冠状面间室压力差>15磅。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Despite advances in total knee arthroplasty (TKA) technology, up to 1 in 5 patients remain dissatisfied. This study sought to evaluate if sensor-guided knee balancing improves postoperative clinical outcomes and patient satisfaction compared to a conventional gap balancing technique. METHODS: We undertook a prospective double-blind randomized controlled trial of patients presenting for elective primary TKA to determine a difference in TKA soft tissue balance between a standard gap balancing (tensiometer) approach compared to augmenting the balance using a sensor-guided device. The sensor-guided experimental group had adjustments made to achieve a balanced knee to within 15 pounds of intercompartmental pressure difference. Secondary outcomes included differences in clinical outcome scores at 6 months and 1 year postoperative, including the Oxford Knee Score and Knee Society Score and patient satisfaction. RESULTS: The sample comprised of 152 patients, 76 controls and 76 experimental sensor-guided cases. Within the control group, 36% (27/76) of knees were unbalanced based on an average coronal plane intercompartmental difference >15 pounds, compared to only 5.3% (4/76) within the experimental group (P < .0001). There were no significant differences in 1-year postoperative flexion, Knee Society Score, or Oxford scores. Overall, TKA patient satisfaction at 1 year was comparable, with 81% of controls and experimental cases reporting they were very satisfied (P = .992). CONCLUSION: Despite the use of the sensor-guided knee balancer device to provide additional quantitative feedback in the evaluation of the soft tissue envelope during TKA, we were unable to demonstrate improved clinical outcomes or patient satisfaction compared to our conventional gap balancing technique.
引用
- Wood TJ, Winemaker MJ, Williams DS, et al. Randomized Controlled Trial of Sensor-Guided Knee Balancing Compared to Standard Balancing Technique in Total Knee Arthroplasty. J Arthroplasty. 2020. PMID: 33041172. doi:10.1016/j.arth.2020.09.025
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。