全膝关节置换术胫骨截骨中摆动锯的软组织穿透
研究在12例尸体膝关节上由3名经验丰富外科医生和3名住院医师行传统截骨板全膝关节置换术,记录胫骨近端截骨时摆动锯超越骨边界的偏移量。结果显示各软组织区域平均第75百分位偏移量为2.8至6.3毫米,经验丰富外科医生偏移量显著低于住院医师,即使经验丰富者摆动锯仍显著超越骨边缘。
要点
- 胫骨截骨时摆动锯在多个软组织区域显著超越骨边缘,平均第75百分位偏移量最高达6.3毫米(外侧副韧带区)。
- 经验丰富外科医生的偏移量显著低于住院医师。
- 即使经验丰富者,摆动锯仍显著超越骨边缘,可造成关节囊和稳定结构直接损伤。
核心结论
- 研究表明,全膝关节置换术胫骨截骨中摆动锯显著超越骨边缘,即使经验丰富者亦如此;尽管神经血管并发症罕见,但手持摆动锯行胫骨截骨可导致关节囊和膝关节周围稳定结构直接损伤。
研究设计与样本
- 在12例尸体膝关节上由3名经验丰富外科医生和3名住院医师行传统截骨板全膝关节置换术(内侧髌旁入路),记录胫骨近端截骨时摆动锯相对胫骨的运动。
主要发现
- 所有病例摆动锯偏移量的平均第75百分位值:内侧副韧带区平均2.8毫米(SD 2.9),后内侧角区平均4.8毫米(SD 5.9),髂胫束区平均3.4毫米(SD 2.0),外侧副韧带区平均6.3毫米(SD 4.8),腘肌腱区平均4.9毫米(SD 5.7),神经血管束区平均6.1毫米(SD 3.9)。
- 经验丰富外科医生的偏移量显著低于住院医师。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIMS: Inadvertent soft tissue damage caused by the oscillating saw during total knee arthroplasty (TKA) occurs when the sawblade passes beyond the bony boundaries into the soft tissue. The primary objective of this study is to assess the risk of inadvertent soft tissue damage during jig-based TKA by evaluating the excursion of the oscillating saw past the bony boundaries. The second objective is the investigation of the relation between this excursion and the surgeon's experience level. METHODS: A conventional jig-based TKA procedure with medial parapatellar approach was performed on 12 cadaveric knees by three experienced surgeons and three residents. During the proximal tibial resection, the motion of the oscillating saw with respect to the tibia was recorded. The distance of the outer point of this cutting portion to the edge of the bone was defined as the excursion of the oscillating saw. The excursion of the sawblade was evaluated in six zones containing the following structures: medial collateral ligament (MCL), posteromedial corner (PMC), iliotibial band (ITB), lateral collateral ligament (LCL), popliteus tendon (PopT), and neurovascular bundle (NVB). RESULTS: The mean 75th percentile value of the excursion of all cases was mean 2.8 mm (SD 2.9) for the MCL zone, mean 4.8 mm (SD 5.9) for the PMC zone, mean 3.4 mm (SD 2.0) for the ITB zone, mean 6.3 mm (SD 4.8) for the LCL zone, mean 4.9 mm (SD 5.7) for the PopT zone, and mean 6.1 mm (SD 3.9) for the NVB zone. Experienced surgeons had a significantly lower excursion than residents. CONCLUSION: This study showed that the oscillating saw significantly passes the edge of the bone during the tibial resection in TKA, even in experienced hands. While reported neurovascular complications in TKA are rare, direct injury to the capsule and stabilizing structures around the knee is a consequence of the use of a hand-held oscillating saw when making the tibial cut. Cite this article: Bone Joint J 2020;102-B(10):1324-1330.
引用
- Herregodts S, Verhaeghe M, Paridaens R, et al. Soft-tissue penetration of the oscillating saw during tibial resection in total knee arthroplasty. Bone Joint J. 2020. PMID: 32993324. doi:10.1302/0301-620X.102B10.BJJ-2019-1602.R2
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。