EOS成像技术对比计算机断层扫描评估青少年特发性脊柱侧凸内固定脊柱椎体旋转方位的准确性

Accuracy of EOS Imaging Technology in Comparison to Computed Tomography in the Assessment of Vertebral Rotational Orientation in Instrumented Spines in Adolescent Idiopathic Scoliosis.
PubMed 文献 · Spine (Phila Pa 1976) · 2021-01-01

该回顾性影像可靠性研究纳入31例接受内固定融合术的青少年特发性脊柱侧凸患者,比较EOS三维重建与CT轴位图像测量椎体旋转的相对椎体旋转差(VRD)。EOS与CT在UIV至顶点(P=0.006)和UIV+1至顶点(P=0.003)的VRD存在显著差异,其余节段无显著差异但Bland-Altman图显示一致性差、方差超出可接受范围。结论认为EOS三维形态分析在内固定节段的VRD测量与CT相比存在显著差异和不可接受方差。

要点

  • EOS与CT在UIV至顶点(P=0.006)和UIV+1至顶点(P=0.003)的VRD存在显著差异。
  • LIV至顶点、LIV+1至顶点、UIV至LIV、UIV+1至LIV+1的VRD无显著差异。
  • Bland-Altman图显示一致性差且方差超出可接受范围,EOS观察者间信度良好。

核心结论

  • EOS三维形态分析在内固定节段的相对椎体旋转差测量与CT相比存在显著差异和不可接受方差。

研究设计与样本

  • 回顾性影像可靠性研究,纳入31例接受内固定融合术的青少年特发性脊柱侧凸患者,术后行脊柱CT扫描。

主要发现

  • EOS与CT测量的相对VRD在UIV至顶点存在显著差异(P=0.006),在UIV+1至顶点存在显著差异(P=0.003)。
  • EOS与CT测量的相对VRD在LIV至顶点(P=0.06)、LIV+1至顶点(P=0.06)、UIV至LIV(P=0.59)和UIV+1至LIV+1(P=0.64)无显著差异。
  • Bland-Altman图显示一致性差且方差超出可接受范围。
  • ICC显示EOS观察者间信度良好,观察者内信度良好至非常好。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

STUDY DESIGN: Retrospective radiographic reliability study. OBJECTIVE: The aim of this study was to assess the validity of EOS 3D imaging technology in the determination of vertebral rotations in the spine of patients with previous instrumentation. SUMMARY OF BACKGROUND DATA: There is a lack of evidence on the accuracy of vertebral rotational measurement using EOS 3D morphological analysis in the instrumented spine. METHODS: A retrospective review of 31 patients with adolescent idiopathic scoliosis (AIS) who underwent instrumented fusion and postoperative computed tomography (CT) scans of the spine was performed. Vertebral rotations of the apex vertebra, the uppermost (UIV) and lowermost (LIV) instrumented vertebra, the noninstrumented vertebra one level cranial to the UIV (UIV + 1) and one level caudal to LIV (LIV + 1) were determined using EOS 3D reconstruction. The vertebral rotation was also measured using reformatted CT axial images. Relative vertebral rotational difference (VRD) were calculated for UIV to apex, UIV + 1 to apex, LIV to apex, LIV + 1 to apex, UIV to LIV and UIV + 1 to LIV + 1. Paired t tests were used to compare the VRD measured using the two different imagining modalities. For values where P > 0.05, the Bland-Altman plot was used to assess the agreement between the measures. Interclass correlation (ICC) was used to determine interobserver and intraobserver reliabilities of EOS and CT measurements. RESULTS: EOS analysis of relative VRD was found to be significantly different from that of CT for UIV to apex (P = 0.006) and UIV + 1 to apex (P = 0.003). No significant differences were found for LIV to apex (P = 0.06), LIV + 1 to apex (P = 0.06), UIV to LIV (P = 0.59) and UIV + 1 to LIV + 1 (P = 0.64). However, Bland-Altman plots showed that agreement was poor, and variance was beyond acceptable. ICC showed good interobserver and good to very good intraobserver reliability for EOS. CONCLUSION: EOS 3D morphological analysis of VRD in the instrumented levels of the spine demonstrated significant difference and unacceptable variance in comparison to CT measurement.Level of Evidence: 4.

来源:Spine (Phila Pa 1976),查看原文。版权归原作者所有。

引用

  1. Jiang F, Simões L, Ouellet J, et al. Accuracy of EOS Imaging Technology in Comparison to Computed Tomography in the Assessment of Vertebral Rotational Orientation in Instrumented Spines in Adolescent Idiopathic Scoliosis. Spine (Phila Pa 1976). 2021. PMID: 33038195. doi:10.1097/BRS.0000000000003739