陶瓷与金属股骨头在5年时的聚乙烯磨损:一项采用放射立体测量分析的随机对照试验

Polyethylene Wear With Ceramic and Metal Femoral Heads at 5 Years: A Randomized Controlled Trial With Radiostereometric Analysis.
PubMed 文献 · J Arthroplasty · 2020-06-23

50例原发性骨关节炎患者按1:1随机接受陶瓷(BIOLOX delta)或金属(CoCr)股骨头非骨水泥全髋关节置换,随访5年重复放射立体测量分析。陶瓷组与金属组早期蠕变分别为0.12 mm与0.08 mm,3个月至5年磨损率分别为0.003 mm/y与0.007 mm/y,两组髋臼杯迁移与临床结果无差异,无翻修。作者结论认为在现代高交联聚乙烯下陶瓷头在早期变形或聚乙烯磨损方面均不优于金属头。

要点

  • 陶瓷组早期蠕变0.12 mm,金属组0.08 mm;3个月至5年磨损率陶瓷组0.003 mm/y、金属组0.007 mm/y。
  • 两组髋臼杯迁移与临床结果无差异,无髋臼杯翻修。
  • 作者结论认为在现代高交联聚乙烯下陶瓷头在早期变形或聚乙烯磨损方面不优于金属头。

核心结论

  • 在现代高交联聚乙烯下,陶瓷头在早期变形或聚乙烯磨损方面均不优于金属头。

研究设计与样本

  • 50例原发性骨关节炎拟行非骨水泥全髋关节置换的患者按1:1随机分为陶瓷(BIOLOX delta)股骨头组或金属(CoCr)股骨头组。
  • 患者随访5年,重复进行放射立体测量分析(术后、14天、3、12、24、60个月)并填写髋关节专用结局问卷。

主要发现

  • 前3个月两组均出现预期内衬蠕变,陶瓷组0.12 mm(标准差0.03),金属组0.08 mm(标准差0.02)。
  • 3个月至5年两组内衬磨损均很小,陶瓷组0.003 mm/y,金属组0.007 mm/y。
  • 两组髋臼杯迁移与临床结果无差异,无髋臼杯翻修。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: A common bearing combination in total hip arthroplasty today is a metal femoral head articulating with polyethylene in the cup. Ceramic heads are thought to be more resistant to third-body damage, and have better wettability and decreased surface roughness, which taken together have been suggested to result in less polyethylene wear. The purpose of this study is to compare the initial creep deformation and follow wear pattern, using radiostereometric analysis, of ceramic and metal femoral heads that articulate with a modern highly cross-linked polyethylene cup liner. METHODS: Fifty patients with primary osteoarthritis and scheduled for an uncemented total hip arthroplasty were randomized 1:1 to either a ceramic (BIOLOX delta) or a metal (CoCr) femoral head. The patients were followed up for 5 years with repeated radiostereometric analysis examinations (postoperatively, then at 14 days, 3, 12, 24, and 60 months), as well as a hip-specific outcome questionnaire. RESULTS: During the first 3 months both groups showed expected creep within the liner of 0.12 mm (standard deviation 0.03) for the ceramic and 0.08 mm (standard deviation 0.02) for the metal heads. Between 3 months and 5 years there was very little wear of the liner in either group, corresponding to 0.003 mm/y for ceramic and 0.007 mm/y for metal heads. There was no difference in cup migration or clinical outcome between the groups and no cups were revised. CONCLUSION: With the introduction of modern highly cross-linked polyethylene, the ceramic head demonstrates no superiority when it comes to either early deformation or polyethylene wear compared with the metal head.